INVESTIGATION OF THE TL CHARACTERISTICS OF COMPOSITE PEAK 5 IN THE GLOW CURVE OF LIF:MG,TI (TLD-100) USING NATURALLY AND FURNACE-COOLED SAMPLES FOLLOWING THE 400°C PRE-IRRADIATION ANNEAL.
Eliyahu, Ilan; Reshes, Galina; Shapiro, Alex; Biderman, Shlomo; Oster, Leonid; Nemirovsky, Dimitry; Sterenberg, Matan; Ginzburg, Dimitry; Horowitz, Yigal Shalom; Herman, Benny; Assor, Yosef.
Afiliação
  • Eliyahu I; Soreq Nuclear Research Center, Yavneh 81800, Israel.
  • Reshes G; Sami Shamoon College of Engineering, Beer Sheva 84100, Israel.
  • Shapiro A; Sami Shamoon College of Engineering, Beer Sheva 84100, Israel.
  • Biderman S; Sami Shamoon College of Engineering, Beer Sheva 84100, Israel.
  • Oster L; Sami Shamoon College of Engineering, Beer Sheva 84100, Israel.
  • Nemirovsky D; Sami Shamoon College of Engineering, Beer Sheva 84100, Israel.
  • Sterenberg M; Ben Gurion University of the Negev, Beer Sheva 84105, Israel.
  • Ginzburg D; Ben Gurion University of the Negev, Beer Sheva 84105, Israel.
  • Horowitz YS; Ben Gurion University of the Negev, Beer Sheva 84105, Israel.
  • Herman B; Sami Shamoon College of Engineering, Beer Sheva 84100, Israel.
  • Assor Y; Nuclear Research Center-Negev, Beer Sheva 84190, Israel.
Radiat Prot Dosimetry ; 196(1-2): 53-59, 2021 Nov 03.
Article em En | MEDLINE | ID: mdl-34463339
ABSTRACT
The cooling rate to room temperature following the 400°C pre-irradiation anneal is known to affect the thermoluminescent properties of LiFMg,Ti (TLD-100) as a result of migration and clustering of defects during the cooling down process. In this investigation the dose response over an extended dose range from 0.01 to 7000 Gy in both naturally cooled and the much slower furnace-cooled samples has been measured. Glow curve deconvolution based on first-order kinetics is employed to extract the dose response of the various glow peaks. Of especial interest is the behaviour of glow peaks 4, 5a and 5 as a function of dose. The idea is to modify the supralinear dose response of peak 5 from 1 to 30 Gy to a linear behaviour. This dose range is important for clinical therapy and a linear dose response is of substantial advantage leading to both improved accuracy and precision.