Your browser doesn't support javascript.
loading
Efficacy and safety of leadless pacemaker: A systematic review, pooled analysis and meta-analysis.
Darlington, Daniel; Brown, Philip; Carvalho, Vanessa; Bourne, Hayley; Mayer, Joseph; Jones, Nathan; Walker, Vincent; Siddiqui, Shoaib; Patwala, Ashish; Kwok, Chun Shing.
Afiliação
  • Darlington D; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Brown P; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Carvalho V; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Bourne H; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Mayer J; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Jones N; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Walker V; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Siddiqui S; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Patwala A; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK.
  • Kwok CS; Department of Cardiology, Royal Stoke University Hospital, Stoke-on-Trent, UK; School of Medicine, Keele University, Keele, UK. Electronic address: shingkwok@doctors.org.uk.
Indian Pacing Electrophysiol J ; 22(2): 77-86, 2022.
Article em En | MEDLINE | ID: mdl-34922032
ABSTRACT

BACKGROUND:

Leadless pacemakers have been designed as an alternative to transvenous systems which avoid some of the complications associated with transvenous devices. We aim to perform a systematic review of the literature to report the safety and efficacy findings of leadless pacemakers.

METHODS:

We searched MEDLINE and EMBASE to identify studies reporting the safety, efficacy and outcomes of patients implanted with a leadless pacemaker. The pooled rate of adverse events was determined and random-effects meta-analysis was performed to compare rates of adverse outcomes for leadless compared to transvenous pacemakers.

RESULTS:

A total of 18 studies were included with 2496 patients implanted with a leadless pacemaker and success rates range between 95.5 and 100%. The device or procedure related death rate was 0.3% while any complication and pericardial tamponade occurred in 3.1% and 1.4% of patients, respectively. Other complications such as pericardial effusion, device dislodgement, device revision, device malfunction, access site complications and infection occurred in less than 1% of patients. Meta-analysis of four studies suggests that there was no difference in hematoma (RR 0.67 95%CI 0.21-2.18, 3 studies), pericardial effusion (RR 0.59 95%CI 0.15-2.25, 3 studies), device dislocation (RR 0.33 95%CI 0.06-1.74, 3 studies), any complication (RR 0.44 95%CI 0.17-1.09, 4 studies) and death (RR 0.45 95%CI 0.15-1.35, 2 studies) comparing patients who received leadless and transvenous pacemakers.

CONCLUSION:

Leadless pacemakers are safe and effective for patients who have an indication for single chamber ventricular pacing and the findings appear to be comparable to transvenous pacemakers.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Prognostic_studies / Systematic_reviews Idioma: En Revista: Indian Pacing Electrophysiol J Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Prognostic_studies / Systematic_reviews Idioma: En Revista: Indian Pacing Electrophysiol J Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Reino Unido