Safety and Performance of POLYTECH Mesmo Breast Implants: A 5-Year Post-market Surveillance Study on 919 Patients.
Aesthet Surg J Open Forum
; 4: ojac011, 2022.
Article
em En
| MEDLINE
| ID: mdl-35415603
Background: In 2007, POLYTECH Health & Aesthetics (POLYTECH, Dieburg, Germany) established an ongoing patient survey to improve the post-market surveillance of silicone gel-filled breast implants based on patient-reported outcomes in the context of the pioneering "Implants of Excellence" (IoE) program. Objectives: To disclose an update on safety and performance outcomes at 5 years for Mesmo breast implants. Methods: Between January 2014 and October 2019, 919 patients (for a total of 1816 implants) who underwent breast augmentation and reconstruction with Mesmo implants were asked to participate in the IoE program. Data were collected by mean of 1320 questionnaires received. A survival analysis assessed the onset of different complications. Results: Eight patients (0.9%) experienced capsular contracture Baker grade III or IV with a cumulative rate at 5 years of 1.2% (95% CI = 0.6-2.4). The proportion of revisional surgery was 0.5% with a 5-year rate of 0.6% (95% CI = 0.2-1.5). Additional adverse events such as hematoma, seroma, malposition, open wounds, and other complications were carefully monitored. Questionnaires showed that 93.9% (95% CI = 92.2-95.4) of the patients were satisfied or very satisfied with their aesthetic results with Mesmo implants. Conclusions: Post-market clinical follow-up revealed that the overall complications rate reported was low. Data demonstrated an excellent safety property on a large cohort of patients. This result allows the rating of Mesmo breast implants as highly competitive and a very safe choice for both surgeons and patients.
Texto completo:
1
Base de dados:
MEDLINE
Tipo de estudo:
Screening_studies
Idioma:
En
Revista:
Aesthet Surg J Open Forum
Ano de publicação:
2022
Tipo de documento:
Article
País de afiliação:
Itália