Your browser doesn't support javascript.
loading
A relaxometry method that emphasizes practicality and availability.
Madore, Bruno; Jerosch-Herold, Michael; Chiou, Jr-Yuan George; Cheng, Cheng-Chieh; Guenette, Jeffrey P; Mihai, Georgeta.
Afiliação
  • Madore B; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • Jerosch-Herold M; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • Chiou JG; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • Cheng CC; Department of Computer Science and Engineering, National Sun Yat-sen University, Kaohsiung, Taiwan.
  • Guenette JP; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • Mihai G; Department of Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Magn Reson Med ; 88(5): 2208-2216, 2022 11.
Article em En | MEDLINE | ID: mdl-35877783
ABSTRACT

PURPOSE:

Although many methods have been proposed to quantitatively map the main MRI parameters (e.g., T1 , T2 , C × M0 ), these methods often involve special sequences not readily available on clinical scanners and/or may require long scan times. In contrast, the proposed method can readily run on most scanners, offer flexible tradeoffs between scan time and image quality, and map MRI parameters jointly to ensure spatial alignment.

METHODS:

The approach is based on the multi-shot spin-echo (SE) EPI sequence. The corresponding signal equation was derived and strategies for solving it were developed. As usual with multi-shot EPI, scan time can readily be traded-off against image quality by adjusting the echo train length. Validation was performed against reference relaxometry methods, in gel phantoms with varying concentrations of gadobutrol and gadoterate meglumine contrast agents. In vivo examples are further presented, from 3 neuroradiology patients.

RESULTS:

Bland-Altman analysis was performed for T2 , as compared to 2D SE, bias was 0.29 ms and the 95% limits of agreement ranged from -1.15 to +1.73 ms. For T1 , compared to inversion-recovery SE (and MOLLI), bias was -20.2 ms (and -14.5 ms) and the limits of agreement ranged from -62.4 to +22.0 ms (and -53.8 to +24.9 ms). The mean relative T1 error between the proposed method and each of the 2 reference methods was similar to that of the reference methods among themselves.

CONCLUSION:

In the constellation of existing relaxometry methods, the proposed method is meant to stand out in terms of its practicality and availability.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos