Your browser doesn't support javascript.
loading
Angiographic quantification of aortic regurgitation following myval octacor implantation; independent core lab adjudication.
Elkoumy, Ahmed; Jose, John; Gunasekaran, Sengottuvelu; Kumar, Asish; Srinivas, B C; Manjunath, C N; Ravindranath, K S; Parekh, Maulik; Chandra, Praveen; Kapoor, Rajneesh; Abdelshafy, Mahmoud; Seth, Ashok; Agrawal, Praveer; Mathur, Atul; Rao, Ravindra Singh; Elzomor, Hesham; Sadanada, K S; Kumar, Vijay; Arsang-Jang, Shahram; Mehrotra, Sanjay; Raghuraman, Bagirath; Khanolkar, Uday; Premchand, Rajendra Kumar; Chopra, Manik; Krishna, Prem; Mehta, Haresh; Gupta, Rahul; Kumar, Viveka; Senguttuvan, Nagendra Boopathy; Baumbach, Andreas; Serruys, Patrick W; Soliman, Osama.
Afiliação
  • Elkoumy A; Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland; Islamic Center of Cardiology, Al-Azhar University, Nasr City, Cairo 11651, Egypt.
  • Jose J; Department of Cardiology, Christian Medical College & Hospital, Vellore 632004, India.
  • Gunasekaran S; Department of Cardiology, Apollo Main Hospital, Greams Road, Chennai 600006, India.
  • Kumar A; Department of Cardiology, Meitra Hospital, Calicut, Kerala 673005, India.
  • Srinivas BC; Sri Jayadeva Institute of Cardiovascular Sciences and Research, Jayanagar, Bengaluru 560041, India.
  • Manjunath CN; Sri Jayadeva Institute of Cardiovascular Sciences and Research, Jayanagar, Bengaluru 560041, India.
  • Ravindranath KS; Sri Jayadeva Institute of Cardiovascular Sciences and Research, Jayanagar, Bengaluru 560041, India.
  • Parekh M; Sir H. N. Reliance Foundation Hospital and Research Centre, Girgaon, Mumbai, Maharashtra 400004, India.
  • Chandra P; Medanta, The Medicity Hospitals, Haryana 122001, India.
  • Kapoor R; Medanta, The Medicity Hospitals, Haryana 122001, India.
  • Abdelshafy M; Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland.
  • Seth A; Fortis Escorts Heart Institute, New Delhi 110025, India.
  • Agrawal P; Fortis Escorts Heart Institute, New Delhi 110025, India.
  • Mathur A; Fortis Escorts Heart Institute, New Delhi 110025, India.
  • Rao RS; Rajasthan Hospital Limited, Jaipur, Rajasthan 302018, India.
  • Elzomor H; Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland.
  • Sadanada KS; Sri Jayadeva Institute of Cardiovascular Sciences and Research, Mysuru, Karnataka 570016, India.
  • Kumar V; Madras Medical Mission Hospital, Chennai, Tamil Nadu 600037, India.
  • Arsang-Jang S; Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland; CÚRAM, SFI Research Centre for Medical Devices, H91 W2TY Galway, Ireland.
  • Mehrotra S; Narayana Institute of Cardiac Sciences, Bengaluru, Karnataka 560099, India.
  • Raghuraman B; Narayana Institute of Cardiac Sciences, Bengaluru, Karnataka 560099, India.
  • Khanolkar U; Narayana Institute of Cardiac Sciences, Bengaluru, Karnataka 560099, India.
  • Premchand RK; Krishna Institute of Medical Sciences (KIMS), Hyderabad, Telangana 500014, India.
  • Chopra M; Narayana Hospital (NH), Ahmedabad, Gujarat 380023, India.
  • Krishna P; P S Govindaswami (PSG), Coimbatore 641004, India.
  • Mehta H; S. L. Raheja Hospital - Fortis, Mumbai, Maharashtra 400016, India.
  • Gupta R; Apollo Hospital, Navi Mumbai, Maharashtra 400614, India.
  • Kumar V; Max Hospital, Saket, New Delhi, Delhi 110017, India.
  • Senguttuvan NB; Sri Ramachandra Hospital (SRMC), Chennai, Tamil Nadu 600116, India.
  • Baumbach A; William Harvey Research Institute, Queen Mary University of London and Barts Heart Centre, London EC1M 6BQ, UK.
  • Serruys PW; Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland; CÚRAM, SFI Research Centre for Medical Devices, H91 W2TY Galway, Ireland.
  • Soliman O; Discipline of Cardiology, Saolta Group, Galway University Hospital, Health Service Executive and CORRIB Core Lab, University of Galway, H91 V4AY Galway, Ireland; CÚRAM, SFI Research Centre for Medical Devices, H91 W2TY Galway, Ireland. Electronic address: osama.soliman@universityofgalway.ie.
Int J Cardiol ; 382: 68-75, 2023 07 01.
Article em En | MEDLINE | ID: mdl-37028710
ABSTRACT

BACKGROUND:

The balloon expandable Myval transcatheter heart valve (THV) showed encouraging results regarding residual aortic regurgitation (AR) from multiple observational studies. The newly designed Myval Octacor has been introduced recently, aiming for a reduction in AR and improved performance.

OBJECTIVES:

The focus of this study is to report the incidence of AR using the validated quantitative Videodensitometry angiography technology (qLVOT-AR%) in the first in human use of the Myval Octacor THV system.

METHODOLOGY:

We report on the first in human use of the Myval Octacor THV system in 125 patients in 18 Indian centres. Independent retrospective analysis of the final aortograms following implantation of the Myval Octacor was performed using the CAAS-A-Valve software. AR is reported as a regurgitation fraction. The previously validated cutoff values have been used to identify ≥moderate AR (RF% >17%), mild (6% < RF% ≤17%), and none or trace AR (RF% ≤ 6%).

RESULTS:

Final aortogram was analysable for 103 patients (84.4%) among the 122 available aortograms. 64 (62%) patients, had tricuspid aortic valve (TAV), 38 (37%) with bicuspid AV (BAV), and one with unicuspid AV. The median absolute RF% was 2% [1, 6], moderate or more AR incidence was 1.9%, mild AR in 20.4%, and none or trace AR in 77.7%. The two cases with RF% >17% were in the BAV group.

CONCLUSION:

The initial results of Myval Octacor using quantitative angiography-derived regurgitation fraction demonstrated a favourable outcome regarding residual AR, possibly due to improved device design. Results must be confirmed in a larger randomised study, including other imaging modalities.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Insuficiência da Valva Aórtica / Estenose da Valva Aórtica / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Clinical_trials / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Int J Cardiol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Egito

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Insuficiência da Valva Aórtica / Estenose da Valva Aórtica / Próteses Valvulares Cardíacas / Substituição da Valva Aórtica Transcateter Tipo de estudo: Clinical_trials / Observational_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Int J Cardiol Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Egito