Construction of a rapid electrochemical biosensor consisting of a nanozyme/aptamer conjugate for waterborne microcystin detection.
Analyst
; 148(11): 2536-2543, 2023 May 30.
Article
em En
| MEDLINE
| ID: mdl-37144330
Microcystin-LR (MC-LR) is a hepatotoxin generated by the excessive proliferation of cyanobacteria, which is a threat to humans and wildlife. Therefore, rapid detection of MC-LR is an important challenge. This study describes a rapid electrochemical biosensor comprising nanozymes and aptamers. Alternating current electrothermal flow (ACEF) significantly reduced the MC-LR detection period to 10 min. We also used MnO2/MC-LR aptamer conjugates to improve the sensitivity to MC-LR detection. Here, MnO2 amplified the electrochemical signal and the aptamer showed high selectivity for MC-LR. Under the optimal conditions, the limit of detection (LOD) and selectivity in freshwater were detected using cyclic voltammetry and differential pulse voltammetry. As a result, an LOD of 3.36 pg mL-1 was observed in the linear concentration range of 10 pg mL-1 to 1 µg mL-1. This study quickly and sensitively detected MC-LR in a situation where it causes serious damage worldwide. In addition, the ACEF technology introduction is the first example of MC-LR detection, suggesting a wide range of possibilities for MC-LR biosensors.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Técnicas Biossensoriais
/
Aptâmeros de Nucleotídeos
Tipo de estudo:
Diagnostic_studies
Limite:
Humans
Idioma:
En
Revista:
Analyst
Ano de publicação:
2023
Tipo de documento:
Article