Accuracy of ultrasound vs. Fourier-domain optic biometry for measuring preoperative axial length in cases of rhegmatogenous retinal detachment.
Jpn J Ophthalmol
; 67(6): 645-651, 2023 Nov.
Article
em En
| MEDLINE
| ID: mdl-37561309
PURPOSE: To identify a method for accurately measuring preoperative axial length (AL) in cases of rhegmatogenous retinal detachment (RRD). STUDY DESIGN: Retrospective study. METHODS: This retrospective study included 83 eyes of 83 patients who underwent vitrectomy for RRD and had both preoperative and postoperative data for AL. Preoperative AL measurements for the affected eye were obtained using ultrasound (aUS-AL) and compared with those for affected and fellow eyes measured using optical biometry (aOB-AL and fOB-AL, respectively). Absolute differences between preoperative aUS-AL, aOB-AL, or fOB-AL measurements and postoperative AL (aPost-AL) were examined. RESULTS: In the 41 eyes without macular detachment, the absolute difference between aOB-AL and aPost-AL (0.06±0.07 mm) was significantly smaller than between aUS-AL and aPost-AL (0.21±0.18 mm) and that between fOB-AL and aPost-AL (0.29±0.35 mm) (P = 0.017 and P < 0.001, respectively). In the 42 eyes with macular detachment, the absolute difference between aOB-AL and aPost-AL (1.22±2.40 mm) was significantly larger than between aUS-AL and aPost-AL (0.24±0.24 mm) and between fOB-AL and aPost-AL (0.35±0.49 mm) (P = 0.006, P = 0.016, respectively). CONCLUSION: The current findings suggest that aOB-AL is more accurate than aUS-AL or fOB-AL in cases of RRD without macular detachment, while aUS-AL or fOB-AL is more accurate than aOB-AL in cases with macular detachment.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Descolamento Retiniano
Tipo de estudo:
Observational_studies
/
Prognostic_studies
Limite:
Humans
Idioma:
En
Revista:
Jpn J Ophthalmol
Ano de publicação:
2023
Tipo de documento:
Article
País de afiliação:
Japão