Your browser doesn't support javascript.
loading
Intra-articular temperature monitoring during radiofrequency ablation in ex-vivo bovine hip joints via Fiber Bragg grating sensors.
Longo, Umile Giuseppe; De Tommasi, Francesca; Salvatore, Giuseppe; Lalli, Alberto; Lo Presti, Daniela; Massaroni, Carlo; Schena, Emiliano.
Afiliação
  • Longo UG; Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo, 200, Roma, 00128, Italy. g.longo@policlinicocampus.it.
  • De Tommasi F; Research Unit of Orthopaedic and Trauma Surgery, Department of Medicine and Surgery, Università Campus Bio-Medico di Roma, Via Alvaro del Portillo, 21, Roma, 00128, Italy. g.longo@policlinicocampus.it.
  • Salvatore G; Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo, 200, Roma, 00128, Italy.
  • Lalli A; Unit of Measurements and Biomedical Instrumentation, Università Campus Bio-Medico di Roma, via Alvaro del Portillo, 200, Trigoria, Rome, 00128, Italy.
  • Lo Presti D; Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo, 200, Roma, 00128, Italy.
  • Massaroni C; Fondazione Policlinico Universitario Campus Bio-Medico, Via Alvaro del Portillo, 200, Roma, 00128, Italy.
  • Schena E; Research Unit of Orthopaedic and Trauma Surgery, Department of Medicine and Surgery, Università Campus Bio-Medico di Roma, Via Alvaro del Portillo, 21, Roma, 00128, Italy.
BMC Musculoskelet Disord ; 24(1): 766, 2023 Sep 28.
Article em En | MEDLINE | ID: mdl-37770871
ABSTRACT

PURPOSE:

Radiofrequency ablation is an increasingly used surgical option for ablation, resection and coagulation of soft tissues in joint arthroscopy. One of the major issues of thermal ablation is the temperature monitoring across the target areas, as cellular mortality is a direct consequence of thermal dosimetry. Temperatures from 45 °C to 50 °C are at risk of damage to chondrocytes. One of the most reliable tools for temperature monitoring is represented by fiber optic sensors, as they allow accurate and real-time temperature measurement via a minimally invasive approach. The aim of this study was to determine, by fiber Bragg grating sensors (FBGs), the safety of radiofrequency ablation in tissue heating applied to ex-vivo bovine hip joints.

METHODS:

Ex vivo bovine hips were subjected to radiofrequency ablation, specifically in the acetabular labrum, for a total of two experiments. The WEREWOLF System (Smith + Nephew, Watford, UK) was employed in high operating mode and in a controlled ablation way. One optical fiber embedding seven FBGs was used to record multipoint temperature variations. Each sensor was 1 mm in length with a distance from edge to edge with each other of 2 mm.

RESULTS:

The maximum variation was recorded in both the tests by the FBG1 (i.e., the closest one to the electrode tip) and was lower than to 2.8 °C. The other sensors (from FBG2 to FBG7) did not record a significant temperature change throughout the duration of the experiment (maximum up to 0.7 °C for FBG7).

CONCLUSIONS:

No significant increase in temperature was observed at any of the seven sites. The sensor nearest to the radiofrequency source exhibited the highest temperature rise, but the variation was only 3 °C. The minimal temperature increase registered at the measurement sites, according to existing literature, is not expected to be cytotoxic. FBGs demonstrate the potential to fulfil the strict requirements for temperature measurements during arthroscopic surgery.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Ablação por Radiofrequência / Articulação do Quadril Limite: Animals / Humans Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Ablação por Radiofrequência / Articulação do Quadril Limite: Animals / Humans Idioma: En Revista: BMC Musculoskelet Disord Assunto da revista: FISIOLOGIA / ORTOPEDIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Itália