Your browser doesn't support javascript.
loading
Dosimetric Characteristics of Radiophotoluminescent Glass Dosimeters for Proton Beams.
Oonsiri, Sornjarod; Kingkaew, Sakda; Vimolnoch, Mananchaya; Chatchumnan, Nichakan; Oonsiri, Puntiwa.
Afiliação
  • Oonsiri S; Department of Radiology, Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Kingkaew S; Department of Radiology, Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Vimolnoch M; Department of Radiology, Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Chatchumnan N; Department of Radiology, Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
  • Oonsiri P; Department of Radiology, Division of Radiation Oncology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
J Med Phys ; 48(3): 238-242, 2023.
Article em En | MEDLINE | ID: mdl-37969142
Purpose: The purpose of the study was to investigate the dosimetric characteristics of radiophotoluminescent glass dosimeters (RGDs) for pencil beam scanning proton therapy. The RGD's end-to-end testing of intensity-modulated proton therapy (IMPT) plans was also evaluated. Materials and Methods: The dosimetric characteristics of the GD-302M type glass dosimeter were studied in terms of uniformity, short-term and long-term reproducibility, stability of the magazine position readout, dose linearity in the range from 0.2 to 20 Gy, energy response in 70-220 MeV, and fading effect. The reference conditions of the spot scanning beam from the Varian ProBeam Compact system were operation at 160 MeV, a 2 cm water-equivalent depth in a solid water phantom, a 10 cm × 10 cm field size at the isocenter, and 2 Gy dose delivery. End-to-end testing of IMPT plans for the head, abdomen, and pelvis was verified using the Alderson Rando phantom. The overall uncertainty analysis was confirmed in this study. Results: The relative response of RGDs for the uniformity test was within 0.95-1.05. The percentages of the coefficients of variation for short-term and long-term reproducibility were 1.16% and 1.50%, respectively. The dose ACE glass dosimetry reader FGD-1000 showed a stable magazine position readout. The dose was found to be linear with R2 = 0.9988. The energy response relative to 160 MeV was approximately within 4.0%. The fading effect was within 2.4%. For the end-to-end test, the difference between the treatment plan and RGD measurement was within 1.0%. The overall uncertainty of the RGD measurement for the proton beam was 4.6%, which covered all energy ranges in this study. Conclusion: The experimental study indicates that the RGDs have the potential to be used in the dosimetry of therapeutic proton beams, including end-to-end dosimetry.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: J Med Phys Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Tailândia

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: J Med Phys Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Tailândia