Multi-peak Light-emitting Diode Curing Units and Polymerization of 3D-printed Crown and Bridge Resin.
Chung, S H; Jiang, E-S; Lim, B-S; Kim, S-H; Chang, J.
Afiliação
  • Chung SH; Shin Hye Chung, DDS, MSD, PhD, Department of Dental Biomaterials Science, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, Republic of Korea.
  • Jiang ES; En-Shi Jiang, DDS, MSD, PhD, Department of Stomatology, Yanbian University and Affiliated Hospital of Yanbian University, Yanji China.
  • Lim BS; Bum-Soon Lim, PhD, Department of Dental Biomaterials Science, School of Dentistry and Dental Research Institute, Seoul National University, Seoul, Republic of Korea.
  • Kim SH; Seong-Hae Kim, Center for Future Dentistry, Seoul National University, Seoul, Republic of Korea.
  • Chang J; *Juhea Chang, DDS, MSD, PhD, National Dental Care Center for Persons with Special Needs, Seoul National University Dental Hospital, Seoul, Republic of Korea.
Oper Dent ; 49(1): 98-109, 2024 Jan 01.
Article em En | MEDLINE | ID: mdl-38058011
PURPOSE: This study aimed to evaluate the substitutive and additive efficacy of multi-peak light-emitting diode (LED) curing units for post-curing of a three-dimensional- (3D-) printed crown and bridge resin. METHODS: A total of 792 disc- and 180 bar-shaped specimens were printed with a crown and bridge resin (NextDent C&B MFH) and post-cured using two LED curing units (VALO Cordless and Bluephase N G4) in fast and standard modes. Conventional post-curing (LC-3D Print Box, Group PC) was compared with LED-only curing (Groups V1, V2, B1, and B2) and LED-combined curing (Groups PV1, PV2, PB1, and PB2) in terms of microhardness, flexural strength, degree of conversion (DC), and CIE L*a*b* color and translucency parameters. Cytotoxicity of the resin eluates was evaluated using the WST-1 assay. Temperature increases on the resin surface were measured with infrared thermography. Data were statistically analyzed using ANOVA and Kruskal-Wallis tests (α=0.05). RESULTS: The microhardness and flexural strength in Groups V1, V2, B1, and DC in all LED-only groups were lower than in other groups (p<0.05). Larger color disparities existed between Group PC and all LED-only groups than between Group PC and the others (p<0.05). There was no significant difference in cytotoxicity among the groups. The temperature increase was lowest in Groups V1 and PV1 during light curing (p<0.05). CONCLUSIONS: Post-curing by multi-peak LED curing units was not as effective as the conventional post-curing device. Additional post-curing by LED curing units did not improve the material properties.