Your browser doesn't support javascript.
loading
Feasibility of Synchrotron-Based Ultra-High Dose Rate (UHDR) Proton Irradiation with Pencil Beam Scanning for FLASH Research.
Yin, Lingshu; Masumi, Umezawa; Ota, Kan; Sforza, Daniel M; Miles, Devin; Rezaee, Mohammad; Wong, John W; Jia, Xun; Li, Heng.
Afiliação
  • Yin L; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Masumi U; Hitachi, Ltd., Research and Development Group, Center for Technology Innovation-Energy, 7-2-1, Omika-chou, Hitachi-shi 319-1292, Ibaraki-ken, Japan.
  • Ota K; Pyramid Technical Consultants, Inc., Boston, MA 02452, USA.
  • Sforza DM; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Miles D; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Rezaee M; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Wong JW; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Jia X; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
  • Li H; Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University School of Medicine, Baltimore, MD 21205, USA.
Cancers (Basel) ; 16(1)2024 Jan 03.
Article em En | MEDLINE | ID: mdl-38201648
ABSTRACT

BACKGROUND:

This study aims to present the feasibility of developing a synchrotron-based proton ultra-high dose rate (UHDR) pencil beam scanning (PBS) system.

METHODS:

The RF extraction power in the synchrotron system was increased to generate 142.4 MeV pulsed proton beams for UHDR irradiation at ~100 nA beam current. The charge per spill was measured using a Faraday cup. The spill length and microscopic time structure of each spill was measured with a 2D strip transmission ion chamber. The measured UHDR beam fluence was used to derive the spot dwell time for pencil beam scanning. Absolute dose distributions at various depths and spot spacings were measured using Gafchromic films in a solid-water phantom.

RESULTS:

For proton UHDR beams at 142.4 MeV, the maximum charge per spill is 4.96 ± 0.10 nC with a maximum spill length of 50 ms. This translates to an average beam current of approximately 100 nA during each spill. Using a 2 × 2 spot delivery pattern, the delivered dose per spill at 5 cm and 13.5 cm depth is 36.3 Gy (726.3 Gy/s) and 56.2 Gy (1124.0 Gy/s), respectively.

CONCLUSIONS:

The synchrotron-based proton therapy system has the capability to deliver pulsed proton UHDR PBS beams. The maximum deliverable dose and field size per pulse are limited by the spill length and extraction charge.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Cancers (Basel) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Cancers (Basel) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos