Linearity and bias of proton density fat fraction across the full dynamic range of 0-100%: a multiplatform, multivendor phantom study using 1.5T and 3T MRI at two sites.
Hu, Houchun H; Chen, Henry Szu-Meng; Hernando, Diego.
Afiliação
  • Hu HH; Department of Radiology, Section of Radiological Science, Anschutz School of Medicine, University of Colorado Denver, Anschutz Medical Campus, Leprino Building, 12401 E 17th Ave, 5th Floor, Mail Stop L954, Aurora, CO, 80045, USA. houchun.hu@cuanschutz.edu.
  • Chen HS; Department of Radiology, Children's Hospital Colorado, Aurora, CO, USA. houchun.hu@cuanschutz.edu.
  • Hernando D; Department of Radiology, Section of Radiological Science, Anschutz School of Medicine, University of Colorado Denver, Anschutz Medical Campus, Leprino Building, 12401 E 17th Ave, 5th Floor, Mail Stop L954, Aurora, CO, 80045, USA.
MAGMA ; 2024 Feb 13.
Article em En | MEDLINE | ID: mdl-38349454
ABSTRACT

OBJECTIVE:

Performance assessments of quantitative determinations of proton density fat fraction (PDFF) have largely focused on the range between 0 and 50%. We evaluate PDFF in a two-site phantom study across the full 0-100% PDFF range. MATERIALS AND

METHODS:

We used commercially available 3D chemical-shift-encoded water-fat MRI sequences from three MRI system vendors at 1.5T and 3T and conducted the study across two sites. A spherical phantom housing 18 vials spanning the full 0-100% PDFF range was used. Data at each site were acquired using default parameters to determine same-day and different-day intra-scanner repeatability, and inter-system and inter-site reproducibility, in addition to linear regression between reference and measured PDFF values.

RESULTS:

Across all systems, results demonstrated strong linearity and minimal bias. For 1.5T systems, a pooled slope of 0.99 with a 95% confidence interval (CI) of 0.981-0.997 and a pooled intercept of 0.61% PDFF with a 95% CI of 0.17-1.04 were obtained. Results for pooled 3T data included a slope of 1.00 (95% CI 0.995-1.005) and an intercept of 0.69% PDFF (95% CI 0.39-0.97). Inter-site and inter-system reproducibility coefficients ranged from 2.9 to 6.2 (in units of PDFF), while intra-scanner same-day and different-day repeatability ranged from 0.6 to 7.8.

DISCUSSION:

PDFF across the 0-100% range can be reliably estimated using current commercial offerings at 1.5T and 3T.