Validation of fluid-structure interaction simulations of the opening phase of phantom mitral heart valves under physiologically inspired conditions.
Christierson, Lea; Frieberg, Petter; Lala, Tania; Töger, Johannes; Liuba, Petru; Revstedt, Johan; Isaksson, Hanna; Hakacova, Nina.
Afiliação
  • Christierson L; Department of Clinical Sciences Lund, Pediatric Heart Center, Skåne University Hospital, Lund University, Lund, Sweden. Address: Barnhjärtcentrum mottagning, Skånes universitetssjukhus, Lasarettsgatan 48, 221 85, Lund, Sweden; Department of Biomedical Engineering, Lund University, Lund, Sweden. Addr
  • Frieberg P; Department of Clinical Sciences Lund, Clinical Physiology, Skåne University Hospital, Lund University, Lund, Sweden. Address: Box 177, 221 00, Lund, Sweden.
  • Lala T; Department of Biomedical Engineering, Lund University, Lund, Sweden. Address: Box 118, 221 00, Lund, Sweden; Department of Clinical Sciences Lund, Clinical Physiology, Skåne University Hospital, Lund University, Lund, Sweden. Address: Box 177, 221 00, Lund, Sweden.
  • Töger J; Department of Clinical Sciences Lund, Clinical Physiology, Skåne University Hospital, Lund University, Lund, Sweden. Address: Box 177, 221 00, Lund, Sweden.
  • Liuba P; Department of Clinical Sciences Lund, Pediatric Heart Center, Skåne University Hospital, Lund University, Lund, Sweden. Address: Barnhjärtcentrum mottagning, Skånes universitetssjukhus, Lasarettsgatan 48, 221 85, Lund, Sweden.
  • Revstedt J; Department of Energy Science, Lund University, Lund, Sweden. Address: Box 118, 221 00, Lund, Sweden.
  • Isaksson H; Department of Biomedical Engineering, Lund University, Lund, Sweden. Address: Box 118, 221 00, Lund, Sweden.
  • Hakacova N; Department of Clinical Sciences Lund, Pediatric Heart Center, Skåne University Hospital, Lund University, Lund, Sweden. Address: Barnhjärtcentrum mottagning, Skånes universitetssjukhus, Lasarettsgatan 48, 221 85, Lund, Sweden.
Comput Biol Med ; 171: 108033, 2024 Mar.
Article em En | MEDLINE | ID: mdl-38430739
ABSTRACT
BACKGROUND AND

OBJECTIVE:

Atrioventricular valve disease is a common cause of heart failure, and successful surgical or interventional outcomes are crucial. Patient-specific fluid-structure interaction (FSI) modeling may provide valuable insights into valve dynamics and guidance of valve repair strategies. However, lack of validation has kept FSI modeling from clinical implementation. Therefore, this study aims to validate FSI simulations against in vitro benchmarking data, based on clinically relevant parameters for evaluating heart valve disease.

METHODS:

An FSI model that mimics the left heart was developed. The domain included a deformable mitral valve of different stiffnesses run with different inlet velocities. Five different cases were simulated and compared to in vitro data based on the pressure difference across the valve, the valve opening, and the velocity in the flow domain.

RESULTS:

The simulations underestimate the pressure difference across the valve by 6.8-14 % compared to catheter measurements. Evaluation of the valve opening showed an underprediction of 5.4-7.3 % when compared to cine MRI, 2D Echo, and 3D Echo data. Additionally, the simulated velocity through the valve showed a 7.9-8.4 % underprediction in relation to Doppler Echo measurements. Qualitative assessment of the velocity profile in the ventricle and the streamlines of the flow in the domain showed good agreement of the flow behavior.

CONCLUSIONS:

Parameters relevant to the diagnosis of heart valve disease estimated by FSI simulations showed good agreement when compared to in vitro benchmarking data, with differences small enough not to affect the grading of heart valve disease. The FSI model is thus deemed good enough for further development toward patient-specific cases.