Commissioning and dosimetric verification of volumetric modulated arc therapy for multiple modalities using electronic portal imaging device-based 3D dosimetry system: a novel approach.
Radiol Phys Technol
; 17(2): 412-424, 2024 Jun.
Article
em En
| MEDLINE
| ID: mdl-38492203
ABSTRACT
The purpose of this study was to validate an electronic portal imaging device (EPID) based 3-dimensional (3D) dosimetry system for the commissioning of volumetric modulated arc therapy (VMAT) delivery for flattening filter (FF) and flattening filter free (FFF) modalities based on test suites developed according to American Association of Physicists in Medicine Task Group 119 (AAPM TG 119) and pre-treatment patient specific quality assurance (PSQA).With ionisation chamber, multiple-point measurement in various planes becomes extremely difficult and time-consuming, necessitating repeated exposure of the plan. The average agreement between measured and planned doses for TG plans is recommended to be within 3%, and both the ionisation chamber and PerFRACTION™ measurement were well within this prescribed limit. Both point dose differences with the planned dose and gamma passing rates are comparable with TG reported multi-institution results. From our study, we found that no significant differences were found between FF and FFF beams for measurements using PerFRACTION™ and ion chamber. Overall, PerFRACTION™ produces acceptable results to be used for commissioning and validating VMAT and for performing PSQA. The findings support the feasibility of integrating PerFRACTION™ into routine quality assurance procedures for VMAT delivery. Further multi-institutional studies are recommended to establish global baseline values and enhance the understanding of PerFRACTION™'s capabilities in diverse clinical settings.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Radiometria
/
Dosagem Radioterapêutica
/
Planejamento da Radioterapia Assistida por Computador
/
Radioterapia de Intensidade Modulada
Limite:
Humans
Idioma:
En
Revista:
Radiol Phys Technol
Assunto da revista:
BIOFISICA
/
RADIOLOGIA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Índia