Your browser doesn't support javascript.
loading
Prospective 1-year results of atrial fibrillation ablation using the pentaspline pulsed field ablation catheter: The initial French experience.
Chaumont, Corentin; McDonnell, Emily; Boveda, Serge; Savoure, Arnaud; Rollin, Anne; Combes, Stephane; Al Hamoud, Raphael; Mandel, Franck; Zeriouh, Sarah; Eltchaninoff, Helene; Maury, Philippe; Anselme, Frederic.
Afiliação
  • Chaumont C; Department of Cardiology, CHU de Rouen, 76000 Rouen, France; University of Rouen Normandie, Inserm EnVI UMR 1096, 76000 Rouen, France.
  • McDonnell E; Department of Cardiology, CHU de Rouen, 76000 Rouen, France.
  • Boveda S; Heart Rhythm Management Department, Clinique Pasteur, 31076 Toulouse, France; Brussels University (VUB), 1050 Brussels, Belgium.
  • Savoure A; Department of Cardiology, CHU de Rouen, 76000 Rouen, France.
  • Rollin A; Department of Cardiology, CHU de Toulouse, 31300 Toulouse, France.
  • Combes S; Heart Rhythm Management Department, Clinique Pasteur, 31076 Toulouse, France.
  • Al Hamoud R; Department of Cardiology, CHU de Rouen, 76000 Rouen, France.
  • Mandel F; Department of Cardiology, CHU de Toulouse, 31300 Toulouse, France.
  • Zeriouh S; Heart Rhythm Management Department, Clinique Pasteur, 31076 Toulouse, France.
  • Eltchaninoff H; Department of Cardiology, CHU de Rouen, 76000 Rouen, France; University of Rouen Normandie, Inserm EnVI UMR 1096, 76000 Rouen, France.
  • Maury P; Department of Cardiology, CHU de Toulouse, 31300 Toulouse, France; I2MC, Inserm UMR 1297, 31432 Toulouse, France.
  • Anselme F; Department of Cardiology, CHU de Rouen, 76000 Rouen, France; University of Rouen Normandie, Inserm EnVI UMR 1096, 76000 Rouen, France. Electronic address: Frederic.anselme@chu-rouen.fr.
Arch Cardiovasc Dis ; 117(4): 249-254, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38494400
ABSTRACT

BACKGROUND:

Pulsed field ablation has recently emerged as an interesting non-thermal energy for atrial fibrillation ablation. At a time of rapid spread of this technology, there is still a lack of prospective real-life data.

AIM:

To describe multicentre prospective safety and 1-year efficacy data in three of the first French centres to use pulsed field ablation.

METHODS:

All consecutive patients undergoing a first pulsed field ablation were included prospectively. The primary outcome was freedom from documented atrial arrhythmia. The safety endpoint was a composite of major adverse events. Univariate and multivariable analyses, including patient and procedural characteristics, were performed to identify factors predictive of recurrence.

RESULTS:

Between May 2021 and June 2022, 311 patients were included (paroxysmal atrial fibrillation in 53%, persistent atrial fibrillation in 35% and long-standing persistent atrial fibrillation in 11%). Additional non-pulmonary vein pulsed field ablation applications were performed in 104/311 patients. One-year freedom from arrhythmia recurrence was 77.6% in the overall population and was significantly higher in patients with paroxysmal atrial fibrillation (88.4%) compared with patients with persistent atrial fibrillation (69.7%; P<0.001) and those with long-standing persistent atrial fibrillation (49.0%; P<0.001). The major complication rate was 2.6% (tamponade in four patients, stroke in two patients and coronary spasm in one patient). Besides the usual predictors of recurrences (left atrium size, CHA2DS2-VASc score, type of atrial fibrillation), the presence of atrial fibrillation at procedure start was independently associated with arrhythmia recurrence (hazard ratio 2.04, 95% confidence interval 1.10-3.77).

CONCLUSION:

In this prospective multicentre real-world study, pulsed field ablation for atrial fibrillation ablation seems to be associated with a good safety profile and rather favourable acute and 1-year success rates.