Your browser doesn't support javascript.
loading
Polyvinyl alcohol hydrogel implant for the treatment of hallux rigidus is associated with a high complication rate and moderate failure rate at short-term follow-up: a systematic review.
Butler, James J; Dhillon, Ravneet; Wingo, Taylor; Lin, Charles C; Samsonov, Alan P; Azam, Mohammad T; Kennedy, John G.
Afiliação
  • Butler JJ; Foot and Ankle Division, Department of Orthopaedic Surgery, NYU Langone Health, 171 Delancey Street, 2nd Floor, New York, NY, 10002, USA. jamesjbutler@rcsi.com.
  • Dhillon R; Royal College of Surgeons in Ireland, Dublin, Ireland.
  • Wingo T; Foot and Ankle Division, Department of Orthopaedic Surgery, NYU Langone Health, 171 Delancey Street, 2nd Floor, New York, NY, 10002, USA.
  • Lin CC; Foot and Ankle Division, Department of Orthopaedic Surgery, NYU Langone Health, 171 Delancey Street, 2nd Floor, New York, NY, 10002, USA.
  • Samsonov AP; Foot and Ankle Division, Department of Orthopaedic Surgery, NYU Langone Health, 171 Delancey Street, 2nd Floor, New York, NY, 10002, USA.
  • Azam MT; Foot and Ankle Division, Department of Orthopaedic Surgery, NYU Langone Health, 171 Delancey Street, 2nd Floor, New York, NY, 10002, USA.
  • Kennedy JG; Foot and Ankle Division, Department of Orthopaedic Surgery, NYU Langone Health, 171 Delancey Street, 2nd Floor, New York, NY, 10002, USA. john.kennedy@nyulangone.org.
Eur J Orthop Surg Traumatol ; 34(4): 1765-1778, 2024 May.
Article em En | MEDLINE | ID: mdl-38554163
ABSTRACT

PURPOSE:

Moderate-to-severe hallux rigidus is a debilitating pathology that is optimally treated with surgical intervention. Arthrodesis produces reliable clinical outcomes but is limited by restriction in 1st metatarsophalangeal joint range of motion. The advent of polyvinyl alcohol hydrogel (PVA) implants have produced early promise based on initial trials, but more recent studies have called into question the efficacy of this procedure. The purpose of this systematic review was to evaluate the clinical and radiological outcomes following the use of PVA for hallux rigidus.

METHODS:

The MEDLINE, EMBASE and Cochrane library databases were systematically reviewed using the preferred reporting items for systematic reviews and meta-analyses guidelines. 18 studies were included.

RESULTS:

In total, 1349 patients (1367 feet) underwent PVA at a weighted mean follow-up of 24.1 ± 11.1 months. There were 168 patients (169 feet) included in the cheilectomy cohort and 322 patients (322 feet) included in the arthrodesis cohort. All 3 cohorts produced comparable improvements in subjective clinical outcomes. Postoperative imaging findings in the PVA cohort included joint space narrowing, peri-implant fluid, peri-implant edema and erosion of the proximal phalanx. The complication rate in the PVA cohort, cheilectomy cohort and arthrodesis cohort was 27.9%, 11.8% and 24.1%, respectively. The failure rates in the PVA cohort, cheilectomy cohort and arthrodesis cohort was 14.8%, 0.3% and 9.0%, respectively.

CONCLUSION:

This systematic review demonstrated that PVA produced a high complication rate (27.9%) together with concerning postoperative imaging findings at short-term follow-up. In addition, a moderate failure rate (14.8%) and secondary surgical procedure rate (9.5%) was noted for the PVA cohort. The findings of this review calls into question the efficacy and safety of PVA for the treatment of hallux rigidus. LEVEL OF EVIDENCE IV.