Your browser doesn't support javascript.
loading
Reliability of the transverse acetabular ligament as a landmark for functional cup anteversion in total hip arthroplasty.
Jassim, Shivan S; Bhatia, Tejasvi; McMahon, Jack R D; Pierrepont, Jim W; McMahon, Stephen J.
Afiliação
  • Jassim SS; Malabar Orthopaedic Clinic, Windsor, VIC, Australia.
  • Bhatia T; Bhatia Orthopaedic Centre, Karnal, India.
  • McMahon JRD; Malabar Orthopaedic Clinic, Windsor, VIC, Australia.
  • Pierrepont JW; Monash University, Melbourne, VIC, Australia.
  • McMahon SJ; University of Sydney, Sydney, Australia.
Hip Int ; 34(5): 608-613, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38572714
ABSTRACT

BACKGROUND:

The transverse acetabular ligament (TAL) can be a useful and reproducible landmark in the orientation of the acetabular cup in total hip arthroplasty (THA). Its role in guiding cup orientation when aiming to implant in a functional anteversion orientation is unclear. The aim of this study was to assess the relationship between the TAL and the planned acetabular cup anteversion when implanted in a function orientation. MATERIAL AND

METHODS:

In a retrospective study the anteversion of the TAL in the contralateral un-replaced hip was measured in CT scans of patients undergoing THA and compared to the functional cup anteversion using the patient-specific spinopelvic parameters. Comparative measurements of the native acetabular version were made from the superior rim to the inferior rim and at 10 mm intervals between the 2, all in reference to the anterior pelvic plane.

RESULTS:

96 hips were measured. The mean TAL anteversion angle was 17.2° ± 4.5°. The mean planned acetabular cup anteversion angle was 26.3° ± 4.7°. Pearson's correlation coefficient of this measurement with the TAL was -0.03 (p = 0.769). There was a significant difference between the planned acetabular cup anteversion and the measurement of the TAL (p < 0.0001).

CONCLUSIONS:

If cups are implanted parallel to the TAL, almost 80% will be >5° different to targeted functional cup version. It should be aimed to implant cups with more anteversion than the TAL indicates.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Artroplastia de Quadril / Prótese de Quadril / Acetábulo / Ligamentos Articulares Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Hip Int Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Austrália

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Artroplastia de Quadril / Prótese de Quadril / Acetábulo / Ligamentos Articulares Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Hip Int Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Austrália