In vivo diffusion MRI of the human heart using a 300 mT/m gradient system.
Afzali, Maryam; Mueller, Lars; Coveney, Sam; Fasano, Fabrizio; Evans, Christopher John; Engel, Maria; Szczepankiewicz, Filip; Teh, Irvin; Dall'Armellina, Erica; Jones, Derek K; Schneider, Jürgen E.
Afiliação
  • Afzali M; Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
  • Mueller L; Cardiff University Brain Research Imaging Centre (CUBRIC), School of Psychology, Cardiff University, Cardiff, UK.
  • Coveney S; Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
  • Fasano F; Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
  • Evans CJ; Siemens Healthcare Ltd, Camberly, UK.
  • Engel M; Siemens Healthcare GmbH, Erlangen, Germany.
  • Szczepankiewicz F; Cardiff University Brain Research Imaging Centre (CUBRIC), School of Psychology, Cardiff University, Cardiff, UK.
  • Teh I; Cardiff University Brain Research Imaging Centre (CUBRIC), School of Psychology, Cardiff University, Cardiff, UK.
  • Dall'Armellina E; Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden.
  • Jones DK; Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
  • Schneider JE; Biomedical Imaging Science Department, Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK.
Magn Reson Med ; 92(3): 1022-1034, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38650395
ABSTRACT

PURPOSE:

This work reports for the first time on the implementation and application of cardiac diffusion-weighted MRI on a Connectom MR scanner with a maximum gradient strength of 300 mT/m. It evaluates the benefits of the increased gradient performance for the investigation of the myocardial microstructure.

METHODS:

Cardiac diffusion-weighted imaging (DWI) experiments were performed on 10 healthy volunteers using a spin-echo sequence with up to second- and third-order motion compensation ( M 2 $$ {M}_2 $$ and M 3 $$ {M}_3 $$ ) and b = 100 , 450 $$ b=100,450 $$ , and 1000 s / m m 2 $$ \mathrm{s}/\mathrm{m}{\mathrm{m}}^2 $$ (twice the b max $$ {b}_{\mathrm{max}} $$ commonly used on clinical scanners). Mean diffusivity (MD), fractional anisotropy (FA), helix angle (HA), and secondary eigenvector angle (E2A) were calculated for b = [100, 450] s / m m 2 $$ \mathrm{s}/\mathrm{m}{\mathrm{m}}^2 $$ and b = [100, 1000] s / m m 2 $$ \mathrm{s}/\mathrm{m}{\mathrm{m}}^2 $$ for both M 2 $$ {M}_2 $$ and M 3 $$ {M}_3 $$ .

RESULTS:

The MD values with M 3 $$ {M}_3 $$ are slightly higher than with M 2 $$ {M}_2 $$ with Δ MD = 0 . 05 ± 0 . 05 [ × 1 0 - 3 mm 2 / s ] ( p = 4 e - 5 ) $$ \Delta \mathrm{MD}=0.05\pm 0.05\kern0.3em \left[\times 1{0}^{-3}\kern0.3em {\mathrm{mm}}^2/\mathrm{s}\right]\kern0.3em \left(p=4e-5\right) $$ for b max = 450 s / mm 2 $$ {b}_{\mathrm{max}}=450\kern0.3em \mathrm{s}/{\mathrm{mm}}^2 $$ and Δ MD = 0 . 03 ± 0 . 03 [ × 1 0 - 3 mm 2 / s ] ( p = 4 e - 4 ) $$ \Delta \mathrm{MD}=0.03\pm 0.03\kern0.3em \left[\times \kern0.3em 1{0}^{-3}\kern0.3em {\mathrm{mm}}^2/\mathrm{s}\right]\kern0.3em \left(p=4e-4\right) $$ for b max = 1000 s / mm 2 $$ {b}_{\mathrm{max}}=1000\kern0.3em \mathrm{s}/{\mathrm{mm}}^2 $$ . A reduction in MD is observed by increasing the b max $$ {b}_{\mathrm{max}} $$ from 450 to 1000 s / mm 2 $$ \mathrm{s}/{\mathrm{mm}}^2 $$ ( Δ MD = 0 . 06 ± 0 . 04 [ × 1 0 - 3 mm 2 / s ] ( p = 1 . 6 e - 9 ) $$ \Delta \mathrm{MD}=0.06\pm 0.04\kern0.3em \left[\times \kern0.3em 1{0}^{-3}\kern0.3em {\mathrm{mm}}^2/\mathrm{s}\right]\kern0.3em \left(p=1.6e-9\right) $$ for M 2 $$ {M}_2 $$ and Δ MD = 0 . 08 ± 0 . 05 [ × 1 0 - 3 mm 2 / s ] ( p = 1 e - 9 ) $$ \Delta \mathrm{MD}=0.08\pm 0.05\kern0.3em \left[\times \kern0.3em 1{0}^{-3}\kern0.3em {\mathrm{mm}}^2/\mathrm{s}\right]\kern0.3em \left(p=1e-9\right) $$ for M 3 $$ {M}_3 $$ ). The difference between FA, E2A, and HA was not significant in different schemes ( p > 0 . 05 $$ p>0.05 $$ ).

CONCLUSION:

This work demonstrates cardiac DWI in vivo with higher b-value and higher order of motion compensated diffusion gradient waveforms than is commonly used. Increasing the motion compensation order from M 2 $$ {M}_2 $$ to M 3 $$ {M}_3 $$ and the maximum b-value from 450 to 1000 s / mm 2 $$ \mathrm{s}/{\mathrm{mm}}^2 $$ affected the MD values but FA and the angular metrics (HA and E2A) remained unchanged. Our work paves the way for cardiac DWI on the next-generation MR scanners with high-performance gradient systems.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imagem de Difusão por Ressonância Magnética / Coração Limite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imagem de Difusão por Ressonância Magnética / Coração Limite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Reino Unido