Endoscopic pressure study integrated system using an ultrathin gastroscope for the functional assessment of the lower esophageal sphincter.
Azuma, Daisuke; Shimamura, Yuto; Inoue, Haruhiro; Tanaka, Hitoshi; Ushikubo, Kei; Yamamoto, Kazuki; Kimoto, Yoshiaki; Okada, Hiroki; Nishikawa, Yohei; Tanaka, Ippei; Onimaru, Manabu.
Afiliação
  • Azuma D; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Shimamura Y; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Inoue H; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Tanaka H; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Ushikubo K; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Yamamoto K; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Kimoto Y; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Okada H; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Nishikawa Y; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Tanaka I; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
  • Onimaru M; Digestive Diseases Center, Showa University Koto Toyosu Hospital, Tokyo, Japan.
Dig Endosc ; 2024 Apr 24.
Article em En | MEDLINE | ID: mdl-38659275
ABSTRACT

OBJECTIVES:

The endoscopic pressure study integrated system (EPSIS) represents an innovative approach for evaluating lower esophageal sphincter function by monitoring intragastric pressure using diagnostic gastroscopes. This study aimed to assess the feasibility and validity of employing ultrathin gastroscopes for EPSIS.

METHODS:

A retrospective analysis was conducted on a database of consecutive patients who underwent EPSIS using both ultrathin and regular gastroscopes between September 2021 and October 2023. The study compared EPSIS parameters between the two gastroscope types to evaluate the correlation of key metrics.

RESULTS:

Thirty patients underwent EPSIS with both ultrathin and regular gastroscopes. Significant positive correlations were observed in the pressure waveform maximum intragastric pressure (mmHg) (ρ = 0.82, P < 0.001) and intragastric pressure gradient (mmHg/s) (ρ = 0.80, P < 0.001) when comparing the two gastroscopes. Maximum intragastric pressure (15.5 [5.3-20.3] vs. 18.5 [3.4-21.6], P < 0.001) and pressure gradient (0.16 [0.013-0.41] vs. 0.24 [0.0039-1.13], P < 0.001), (median [range]) were significantly lower with ultrathin gastroscopes.

CONCLUSIONS:

This study establishes that EPSIS parameters obtained with an ultrathin gastroscope exhibit a significant correlation with those obtained using a regular gastroscope, with each EPSIS parameter consistently lower. These findings support the viability of EPSIS for ultrathin gastroscopy and highlight its potential as a diagnostic tool for assessing lower esophageal sphincter function.