Validation of echo planar imaging based diffusion-weighted magnetic resonance imaging on a 0.35 T MR-Linac.
Wallimann, Philipp; Piccirelli, Marco; Nowakowska, Sylwia; Armstrong, Tess; Mayinger, Michael; Boss, Andreas; Bink, Andrea; Guckenberger, Matthias; Tanadini-Lang, Stephanie; Andratschke, Nicolaus; Pouymayou, Bertrand.
Afiliação
  • Wallimann P; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Piccirelli M; Department of Neuroradiology, Clinical Neuroscience Center, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Nowakowska S; Institute for Diagnostic and Interventional Radiology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Armstrong T; ViewRay Inc., 2 Thermo Fisher Way, Oakwood Village, OH 44146, USA.
  • Mayinger M; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Boss A; Institute for Diagnostic and Interventional Radiology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Bink A; Department of Neuroradiology, Clinical Neuroscience Center, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Guckenberger M; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Tanadini-Lang S; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Andratschke N; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
  • Pouymayou B; Department of Radiation Oncology, University Hospital Zurich and University of Zurich, Zurich, Switzerland.
Phys Imaging Radiat Oncol ; 30: 100579, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38707628
ABSTRACT
Background and

Purpose:

The feasibility of acquiring diffusion-weighted imaging (DWI) images on an MR-Linac for quantitative response assessment during radiotherapy was explored. DWI data obtained with a Spin Echo Echo Planar Imaging sequence adapted for a 0.35 T MR-Linac were examined and compared with DWI data from a conventional 3 T scanner. Materials and

Methods:

Apparent diffusion coefficient (ADC) measurements and a distortion correction technique were investigated using DWI-calibrated phantoms and in the brains of seven volunteers. All DWI utilized two phase-encoding directions for distortion correction and off-resonance field estimation. ADC maps in the brain were analyzed for automatically segmented normal tissues.

Results:

Phantom ADC measurements on the MR-Linac were within a 3 % margin of those recorded by the 3 T scanner. The maximum distortion observed in the phantom was 2.0 mm prior to correction and 1.1 mm post-correction on the MR-Linac, compared to 6.0 mm before correction and 3.6 mm after correction at 3 T. In vivo, the average ADC values for gray and white matter exhibited variations of 14 % and 4 %, respectively, for different selections of b-values on the MR-Linac. Distortions in brain images before correction, estimated through the off-resonance field, reached 2.7 mm on the MR-Linac and 12 mm at 3 T.

Conclusion:

Accurate ADC measurements are achievable on a 0.35 T MR-Linac, both in phantom and in vivo. The selection of b-values significantly influences ADC values in vivo. DWI on the MR-Linac demonstrated lower distortion levels, with a maximum distortion reduced to 1.1 mm after correction.