Histotripsy and Catheter-Directed Lytic: Efficacy in Highly Retracted Porcine Clots In Vitro.
Yang, Shumeng; Zemzemi, Chadi; Escudero, Daniel Suarez; Vela, Deborah C; Haworth, Kevin J; Holland, Christy K.
Afiliação
  • Yang S; Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH, USA. Electronic address: yang2s5@mail.uc.edu.
  • Zemzemi C; Department of Internal Medicine, University of Cincinnati, Cincinnati, OH, USA.
  • Escudero DS; Department of Internal Medicine, University of Cincinnati, Cincinnati, OH, USA.
  • Vela DC; Cardiovascular Pathology, Texas Heart Institute, Houston, TX, USA.
  • Haworth KJ; Department of Internal Medicine, University of Cincinnati, Cincinnati, OH, USA; Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH, USA.
  • Holland CK; Department of Internal Medicine, University of Cincinnati, Cincinnati, OH, USA; Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH, USA.
Ultrasound Med Biol ; 50(8): 1167-1177, 2024 Aug.
Article em En | MEDLINE | ID: mdl-38777639
ABSTRACT

OBJECTIVE:

Standard treatment for deep vein thrombosis (DVT) involves catheter-directed anticoagulants or thrombolytics, but the chronic thrombi present in many DVT cases are often resistant to this therapy. Histotripsy has been found to be a promising adjuvant treatment, using the mechanical action of cavitating bubble clouds to enhance thrombolytic activity. The objective of this study was to determine if histotripsy enhanced recombinant tissue plasminogen activator (rt-PA) thrombolysis in highly retracted porcine clots in vitro in a flow model of occlusive DVT.

METHODS:

Highly retracted porcine whole blood clots were treated for 1 h with either catheter-directed saline (negative control), rt-PA (lytic control), histotripsy, DEFINITY and histotripsy or the combination of rt-PA and histotripsy with or without DEFINITY. Five-cycle, 1.5 MHz histotripsy pulses with a peak negative pressure of 33.2 MPa and pulse repetition frequency of 40 Hz were applied along the clot. B-Mode and passive cavitation images were acquired during histotripsy insonation to monitor bubble activity.

RESULTS:

Clots subjected to histotripsy with and without rt-PA exhibited greater thrombolytic efficacy than controls (7.0% flow recovery or lower), and histotripsy with rt-PA was more efficacious than histotripsy with saline (86.1 ± 10.2% compared with 61.7 ± 19.8% flow recovery). The addition of DEFINITY to histotripsy with or without rt-PA did not enhance either thrombolytic efficacy or cavitation dose. Cavitation dose generally did not correlate with thrombolytic efficacy.

CONCLUSION:

Enhancement of thrombolytic efficacy was achieved using histotripsy, with and without catheter-directed rt-PA, in the presence of physiologic flow. This suggests these treatments may be effective as therapy for DVT.