Impact of scatter radiation on spectral quantification performance of first- and second-generation dual-layer spectral computed tomography.
Salazar, Edgar; Liu, Leening P; Perkins, Amy E; Halliburton, Sandra S; Shapira, Nadav; Litt, Harold I; Noël, Peter B.
Afiliação
  • Salazar E; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Liu LP; Department of Engineering and Architecture, Universidad Privada Boliviana, La Paz, Bolivia.
  • Perkins AE; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Halliburton SS; Philips Healthcare, Orange Village, Ohio, USA.
  • Shapira N; Philips Healthcare, Orange Village, Ohio, USA.
  • Litt HI; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Noël PB; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
J Appl Clin Med Phys ; 25(7): e14383, 2024 Jul.
Article em En | MEDLINE | ID: mdl-38801204
ABSTRACT

OBJECTIVE:

To assess the impact of scatter radiation on quantitative performance of first and second-generation dual-layer spectral computed tomography (DLCT) systems.

METHOD:

A phantom with two iodine inserts (1 and 2 mg/mL) configured to intentionally introduce high scattering conditions was scanned with a first- and second-generation DLCT. Collimation widths (maximum of 4 cm for first generation and 8 cm for second generation) and radiation dose levels were varied. To evaluate the performance of both systems, the mean CT numbers of virtual monoenergetic images (MonoEs) at different energies were calculated and compared to expected values. MonoEs at 50  versus 150 keV were plotted to assess material characterization of both DLCTs. Additionally, iodine concentrations were determined, plotted, and compared against expected values. For each experimental scenario, absolute errors were reported.

RESULTS:

An experimental setup, including a phantom design, was successfully implemented to simulate high scatter radiation imaging conditions. Both CT scanners illustrated high spectral accuracy for small collimation widths (1 and 2 cm). With increased collimation (4 cm), the second-generation DLCT outperformed the earlier DLCT system. Further, the spectral performance of the second-generation DLCT at an 8 cm collimation width was comparable to a 4 cm collimation on the first-generation DLCT. A comparison of the absolute errors between both systems at lower energy MonoEs illustrates that, for the same acquisition parameters, the second-generation DLCT generated results with decreased errors. Similarly, the maximum error in iodine quantification was less with second-generation DLCT (0.45  and 0.33 mg/mL for the first and second-generation DLCT, respectively).

CONCLUSION:

The implementation of a two-dimensional anti-scatter grid in the second-generation DLCT improves the spectral quantification performance. In the clinical routine, this improvement may enable additional clinical benefits, for example, in lung imaging.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espalhamento de Radiação / Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X / Imagens de Fantasmas Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espalhamento de Radiação / Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X / Imagens de Fantasmas Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos