Inoculum Size and False-Positive Detection of NDM- and OXA-48-Type Carbapenemases Using Two Multiplex Lateral Flow Assays.
Diagnostics (Basel)
; 14(12)2024 Jun 17.
Article
em En
| MEDLINE
| ID: mdl-38928689
ABSTRACT
The NG-Test CARBA 5 and Carbapenem-resistant K.N.I.V.O. Detection K-Set are lateral flow assays (LFAs) that rapidly detect five carbapenemases (KPC, NDM, IMP, VIM and OXA-48-like). We evaluated the effect of inoculum size on the performance of these two assays using 27 Enterobacterales isolates. Whole-genome sequencing (WGS) was used as the reference method. Using the NG-Test CARBA 5, eight Serratia spp. and six M. morganii isolates showed false-positive NDM results with a high inoculum. Using the Carbapenem-resistant K.N.I.V.O. Detection K-Set, eight M. morganii, four Serratia spp. and one K. pneumoniae isolates showed false-positive NDM and/or OXA-48-like bands at large inoculum sizes, while the other two M. morganii isolates demonstrated false-positive NDM and OXA-48-like results at all inoculum sizes. The false-positive bands varied in intensity. WGS confirmed that no carbapenemase gene was present. No protein sequence with a ≥50% identity to NDM or OXA-48-like enzymes was found. This study emphasizes the importance of assessing inoculum size in the diagnostic evaluation of LFAs.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Revista:
Diagnostics (Basel)
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
China