Combined Rotating Ultra-High-Resolution Spectral Domain OCT and Scheimpflug Imaging for In Vivo Corneal Optical Biopsy.
Ambrósio, Renato; Esporcatte, Louise Pellegrino G; de Carvalho, Karolyna Andrade; Salomão, Marcella Q; Pereira-Souza, Amanda Luiza; Lopes, Bernardo T; Machado, Aydano P; Marschall, Sebastian.
Afiliação
  • Ambrósio R; Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Rio de Janeiro 20520-050, Brazil.
  • Esporcatte LPG; Instituto de Olhos Renato Ambrósio, Rio de Janeiro 20520-050, Brazil.
  • de Carvalho KA; Department of Ophthalmology, Federal University of São Paulo, São Paulo 04021-001, Brazil.
  • Salomão MQ; Brazilian Artificial Intelligence Networking in Medicine-BrAIN, Rio de Janeiro 20520-050, Brazil.
  • Pereira-Souza AL; Brazilian Artificial Intelligence Networking in Medicine-BrAIN, Maceió 57072-900, Brazil.
  • Lopes BT; Department of Ophthalmology, Federal University the State of Rio de Janeiro (UNIRIO), Rio de Janeiro 20520-050, Brazil.
  • Machado AP; Rio de Janeiro Corneal Tomography and Biomechanics Study Group, Rio de Janeiro 20520-050, Brazil.
  • Marschall S; Instituto de Olhos Renato Ambrósio, Rio de Janeiro 20520-050, Brazil.
Diagnostics (Basel) ; 14(13)2024 Jul 08.
Article em En | MEDLINE | ID: mdl-39001345
ABSTRACT

PURPOSE:

This article introduces the Pentacam® Cornea OCT (optical coherence tomography). This advanced corneal imaging system combines rotating ultra-high-resolution spectral domain OCT with sub- 2-micron axial resolution and Scheimpflug photography. The purpose of this study is to present the first experience with the instrument and its potential for corneal diagnostics, including optical biopsy.

METHODS:

In this prospective study, the Pentacam® Cornea OCT was used to image the corneas of seven patients. The novel wide-angle pericentric scan system enables optimal OCT imaging performance for the corneal layer structure over the entire width of the cornea, including the limbal regions. A detailed analysis of the resulting images assessed the synergism between the OCT and Scheimpflug photography.

RESULTS:

The Pentacam® Cornea OCT demonstrated significantly improved image resolution and ability to individualize corneal layers with high quality. There is a synergism between the OCT high-definition signal to individualize details on the cornea and Scheimpflug photography to detect and quantify corneal scattering. The noncontact exam was proven safe, user-friendly, and effective for enabling optical biopsy.

CONCLUSIONS:

Pentacam® Cornea OCT is an advancement in corneal imaging technology. The ultra-high-resolution spectral domain OCT and Scheimpflug photography provide unprecedented detail and resolution, enabling optical biopsy and improving the understanding of corneal pathology. Further studies are necessary to compare and analyze the tomographic reconstructions of the cornea with the different wavelengths, which may provide helpful information for diagnosing and managing corneal diseases.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Diagnostics (Basel) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Base de dados: MEDLINE Idioma: En Revista: Diagnostics (Basel) Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Brasil