Comparison of the spectral performance between two dual-source CT systems on low-energy virtual monoenergetic images: A phantom study.
Greffier, Joël; Faby, Sebastian; Pastor, Maxime; Frandon, Julien; Erath, Julien; Beregi, Jean-Paul; Dabli, Djamel.
Afiliação
  • Greffier J; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France. Electronic address: joel.greffier@chu-nimes.fr.
  • Faby S; Department of Computed Tomography, Siemens Healthineers AG, Siemensstr. 3, 91301 Forchheim, Germany.
  • Pastor M; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
  • Frandon J; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
  • Erath J; Department of Computed Tomography, Siemens Healthineers AG, Siemensstr. 3, 91301 Forchheim, Germany.
  • Beregi JP; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
  • Dabli D; IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
Phys Med ; 124: 103429, 2024 Aug.
Article em En | MEDLINE | ID: mdl-39024963
ABSTRACT

PURPOSE:

To compare the spectral performance of two different DSCT (DSCT-Pulse and DSCT-Force) on virtual monoenergetic images (VMIs) at low energy levels.

METHODS:

An image quality phantom was scanned on the two DSCTs at three dose levels 11/6/1.8 mGy. Level 3 of an advanced modeled iterative reconstruction algorithm was used. Noise power spectrum and task-based transfer function were computed on VMIs from 40 to 70 keV to assess noise magnitude and noise texture (fav) and spatial resolution (f50). A detectability index (d') was computed to assess the detection of one contrast-enhanced abdominal lesion as a function of the keV level used.

RESULTS:

For all dose levels and all energy levels, noise magnitude was significantly higher (p < 0.05) with DSCT-Pulse than with DSCT-Force (12.6 ± 2.7 % at 1.8 mGy, 9.1 ± 2.9 % at 6 mGy and 4.0 ± 2.7 % at 11 mGy). For all energy levels, fav values were significantly higher (p < 0.05) with DSCT-Pulse than with DSCT-Force at 1.8 mGy (4.8 ± 3.9 %) and at 6 mGy (5.5 ± 2.5 %) but similar at 11 mGy (0.2 ± 3.6 %; p = 0.518). For all energy levels, f50 values were significantly higher with DSCT-Pulse than with DSCT-Force (12.7 ± 5.6 % at 1.8 mGy, 17.9 ± 4.5 % at 6 mGy and 13.1 ± 2.6 % at 11 mGy). For all keV, similar d' values were found with both DSCT-Force and DSCT-Pulse at 11 mGy (-1.0 ± 3.1 %; p = 0.084). For other dose levels, d' values were significantly lower with DSCT-Pulse than with DSCT-Force (9.1 ± 3.2 % at 1.8 mGy and -6.3 ± 3.9 % at 6 mGy).

CONCLUSION:

Compared with the DSCT-Force, the DSCT-Pulse improved noise texture and spatial resolution, but noise magnitude was slightly higher and detectability slightly lower, particularly when the dose level was reduced.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X / Imagens de Fantasmas Limite: Humans Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X / Imagens de Fantasmas Limite: Humans Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2024 Tipo de documento: Article