Excimer laser surgery of the cornea: qualitative and quantitative aspects of photoablation according to the energy density.
J Cataract Refract Surg ; 12(1): 27-33, 1986 Jan.
Article em En | MEDLINE | ID: mdl-3958947
Using an excimer laser working in the far ultraviolet and cutting fresh corneas of human cadaver eyes, we were able to demonstrate that there is an ideal fluence (joule/cm2/pulse) that allows a maximum tissue ablation depth per shot. Above the threshold of 1 J/cm2/pulse the depth of tissue ablated does not increase and both thermal and mechanical damage appears at the cut edges.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Córnea / Terapia a Laser Tipo de estudo: Qualitative_research Limite: Humans Idioma: En Revista: J Cataract Refract Surg Assunto da revista: OFTALMOLOGIA Ano de publicação: 1986 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Córnea / Terapia a Laser Tipo de estudo: Qualitative_research Limite: Humans Idioma: En Revista: J Cataract Refract Surg Assunto da revista: OFTALMOLOGIA Ano de publicação: 1986 Tipo de documento: Article