Microdensitometry using a slide projector.
Med Phys ; 12(4): 480-1, 1985.
Article em En | MEDLINE | ID: mdl-4033594
ABSTRACT
We have evaluated a simple microdensitometer concept by Stanton et al. for measurements of dose distributions with high spatial resolution. A magnified image of the film is created by a slide projector and its light intensity measured with a photomultiplier with a small aperture. The projector is placed on a movable stage to allow scanning of the film. A resolution of 17 micron (full width at half maximum of the line spread function) was achieved in the film plane.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Absorciometria de Fóton / Óptica e Fotônica Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1985 Tipo de documento: Article