Laser anemometry measurements of pulsatile flow past aortic valve prostheses.
J Biomech ; 16(10): 865-73, 1983.
Article em En | MEDLINE | ID: mdl-6643525
ABSTRACT
Experimental results are presented on physiological pulsatile flow past caged ball and tilting disc aortic valve prostheses mounted in an axisymmetric chamber incorporated in a mock circulatory system. The measurements of velocity profiles and turbulent normal stresses during several times in a cardiac cycle were obtained using laser-Doppler anemometry. Our results show that with increased angle of opening for the tilting disc valves, a large but locally confined vortex is observed along the wall in the minor flow region throughout most of the cardiac cycle. The turbulent normal stresses measured downstream to the tilting disc in the minor flow region parallel to the tilt axis were found to be larger than those measured downstream to the caged ball valves. Comparison of measurements with steady flow at flow rates comparable to peak pulsatile flow rate show that the turbulent normal stresses are larger by a factor of two in pulsatile flow with a frequency of 1.2 Hz.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Próteses Valvulares Cardíacas Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: J Biomech Ano de publicação: 1983 Tipo de documento: Article