ABSTRACT
In this paper, laser-Doppler anemometry measurement of steady flow development in a model human aorta has been reported. Studies were made with uniform entry flow at the root of the aorta and our measurements showed the establishment of a pair of Dean vortices in the mid-arch region. Subsequently, the nature of flow development past centrally occluding caged ball valves in the model aorta was investigated. Our studies showed that in the ascending aorta, an asymmetric velocity profile is obtained with larger velocity gradients towards the inner wall of tertiary curvature (anatomically the left lateral wall) with centrally occluding valves. The peripheral flow past these valves prevented the development of Dean vortices in the mid-arch region. The caged ball valves at the root of the aorta had no discernible effect on the velocity profiles in the brachio-cephalic artery.