Liquid chromatographic determination of the adenosine receptor agonist CGS 21680 in blood using on-line solid-phase extraction on a phenylboronic acid support and fluorescence detection.
J Chromatogr B Biomed Appl
; 658(1): 207-12, 1994 Aug 05.
Article
em En
| MEDLINE
| ID: mdl-7952127
ABSTRACT
An analytical method is described for the selective determination of A1 or A2 adenosine receptor agonists in blood. By implementing solid-phase extraction using immobilized-phenylboronic acid (PBA) in sample pretreatment, all adenosine derivatives are retained via their intact cis-diol group. On-line desorption of the analytes from the PBA support to the C18 analytical column is performed by injection of a small plug of perchloric acid. Fluorescence and UV detection are employed for the different adenosine derivatives. The method is applied to the determination of 2-[p-(2-carboxyethyl)phenylethylamino]-5'-N- ethylcarboxyamidoadenosine (CGS 21680, I) in blood using fluorescence detection. The only off-line sample handling step is the extraction of blood with ethyl acetate and subsequent evaporation of the extraction solvent. The detection limit of the method was 0.25 ng (signal-to-noise ratio 31) and the determination limit for I in blood (pretreatment of 100 microliters) was 5 ng/ml. The method was validated and used to study the pharmacokinetics of I in rats.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Fenetilaminas
/
Ácidos Borônicos
/
Adenosina
/
Cromatografia Líquida de Alta Pressão
/
Agonistas do Receptor Purinérgico P1
Tipo de estudo:
Diagnostic_studies
Limite:
Animals
Idioma:
En
Revista:
J Chromatogr B Biomed Appl
Ano de publicação:
1994
Tipo de documento:
Article
País de afiliação:
Holanda