Accuracy of the urinary albumin titrator stick 'Micral-Test' in kidney-disease patients.
Minetti, E E; Cozzi, M G; Granata, S; Guidi, E.
Afiliação
  • Minetti EE; Centro di Ricerca Clinica in Nefrologia e Ipertensione Arteriosa-Unità Operativa di Nefrologia, Dialisi e Terapia del Trapianto Renale, Ospedale Niguarda Ca' Granda, Milano, Italy.
Nephrol Dial Transplant ; 12(1): 78-80, 1997 Jan.
Article em En | MEDLINE | ID: mdl-9027777
ABSTRACT

BACKGROUND:

A quick, accurate, and easy measurement of microalbuminuria can be useful for the management of many patients. The Micral-Test stick (Boehringer Mannheim, Germany) gives a semiquantitative estimation of urinary albumin concentration at discrete readings of 0, 10, 20, 50 or 100 mg/l. The purpose of this study was to test its accuracy.

METHODS:

From 46 non-diabetic patients, 491 urinary samples were analysed both with Micral-Test and with immunochemical nephelometry.

RESULTS:

Of 491 samples, 308 were from non-proteinuric patients (urinary albumin < 300 mg/day). In these patients the correlation coefficient of nephelometric values versus the stick readings was 0.79 120/123 samples from non-microalbuminuric (< or = 30 mg/24 h) and 164/185 from microalbuminuric patients were correctly identified by the stick, giving an 89% sensitivity and a 98% specificity. The readings around the threshold for microalbuminuria (20 and 50 mg/l patches) were the most reliable ones. Analysis of the correct/uncorrect readings' ratio for every patch in the 245 samples in the 0-150 mg/l range shows a relative uniformity (chi2 = 8.5, P = 0.07), while analysis of the over/correct/underreadings for the 10, 20 and 50 patches, shows that incorrect responses tend to be underestimations for the 10 patch and overestimations for the 20 and 50 mg/l patches (chi2 = 10.5, P = 0.03).

CONCLUSIONS:

The Micral-Test stick is useful for screening, but less reliable for follow-up, unless considerable changes in albumin excretion are expected.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Ensaio de Imunoadsorção Enzimática / Albuminúria / Nefropatias Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Nephrol Dial Transplant Assunto da revista: NEFROLOGIA / TRANSPLANTE Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Itália
Buscar no Google
Base de dados: MEDLINE Assunto principal: Ensaio de Imunoadsorção Enzimática / Albuminúria / Nefropatias Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Nephrol Dial Transplant Assunto da revista: NEFROLOGIA / TRANSPLANTE Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Itália