[Is Ambu ventilation of newborn infants a simple question of finger-touch?]. / La ventilation à l'Ambu chez le nouveau-né: une simple question de doigté?
Ganga-Zandzou, P S; Diependaele, J F; Storme, L; Riou, Y; Klosowski, S; Rakza, T; Logier, R; Lequien, P.
Afiliação
  • Ganga-Zandzou PS; Service de médecine néonatale, CHFU de Lille, hôpital Jeanne-de-Flandre, France.
Arch Pediatr ; 3(12): 1270-2, 1996 Dec.
Article em Fr | MEDLINE | ID: mdl-9033794
ABSTRACT
Insufflation pressures were measured during manual ventilation using a neonatal rebreathing bag (Ambu on a manikin. Maximal insufflation pressures were greater than that published or given by the manufacturer, theoretically limited to 30 cm of water at open valve, and that whatever the number of fingers used for the compression of the bag. These results indicate that Ambu ventilation, often mandatory for newborn resuscitation, does not simply rely upon the finger-touch of the operator and that it always has a risk of baro and/or volotraumatism.