Your browser doesn't support javascript.
loading
Prolonged hemodynamic stability during arteriovenous carbon dioxide removal for severe respiratory failure.
Brunston, R L; Tao, W; Bidani, A; Alpard, S K; Traber, D L; Zwischenberger, J B.
Afiliação
  • Brunston RL; Department of Surgery, University of Texas Medical Branch, Galveston 77550-0528, USA.
J Thorac Cardiovasc Surg ; 114(6): 1107-14, 1997 Dec.
Article em En | MEDLINE | ID: mdl-9434706
OBJECTIVE: The effects of prolonged arteriovenous carbon dioxide removal on hemodynamics during severe respiratory failure were evaluated in adult sheep with severe smoke inhalation injury. METHODS: Adult female sheep (n = 6,33.8 +/- 5.2 kg) were subjected to intratracheal cotton severe smoke insufflation to a mean carboxyhemoglobin level of 83% +/- 3%. Twenty-four hours after injury, a low-resistance 2.5 m2 membrane oxygenator was placed in a carotid-to-jugular pumpless arteriovenous shunt at unrestricted flow to allow complete carbon dioxide removal and reductions in ventilator support. Animals remained conscious, and heart rate, cardiac output, mean arterial pressure, and pulmonary arterial pressure were measured at baseline, after injury, and daily during support with the arteriovenous carbon dioxide removal circuit for 7 days. RESULTS: All animals survived the study period. Carbon dioxide removal ranged from 99.7 +/- 13.7 to 152.2 +/- 16.2 ml/min, and five (83%) of the six animals were successfully weaned from the ventilator before day 7. During full support with the arteriovenous carbon dioxide removal circuit, shunt flow ranged from 1.24 +/- 0.06 to 1.43 +/- 0.08 L/min and accounted for 20.1% +/- 1.4% to 25.9% +/- 2.4% of cardiac output. No statistically significant changes in heart rate, cardiac output, mean arterial pressure, or pulmonary artery pressure were demonstrated over the study course despite the extracorporeal shunt flow. CONCLUSIONS: Arteriovenous carbon dioxide removal as a simplified means of extracorporeal gas exchange support is relatively safe without adverse hemodynamic effects or complications.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Síndrome do Desconforto Respiratório / Dióxido de Carbono / Oxigenação por Membrana Extracorpórea / Lesão por Inalação de Fumaça / Hemodinâmica Limite: Animals Idioma: En Revista: J Thorac Cardiovasc Surg Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Síndrome do Desconforto Respiratório / Dióxido de Carbono / Oxigenação por Membrana Extracorpórea / Lesão por Inalação de Fumaça / Hemodinâmica Limite: Animals Idioma: En Revista: J Thorac Cardiovasc Surg Ano de publicação: 1997 Tipo de documento: Article País de afiliação: Estados Unidos