Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros

Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Braz J Cardiovasc Surg ; 39(1): e20230110, 2023 11 09.
Artigo em Inglês | MEDLINE | ID: mdl-37947184

RESUMO

OBJECTIVE: To define a reference chart comparing pressure drop vs. flow generated by a set of arterial cannulae currently utilized in cardiopulmonary bypass conditions in pediatric surgery. METHODS: Cannulae from two manufacturers were selected considering their design and outer and inner diameters. Cannula performance was evaluated in terms of pressure drop vs. flow during simulated cardiopulmonary bypass conditions. The experimental circuits consisted of a Jostra HL-20 roller pump, a Quadrox-i pediatric oxygenator (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set. The circuit was primed with lactated Ringer's solution only (first condition) and with human packed red blood cells added (second condition) to achieve a hematocrit of 30%. Cannula sizes 8 to 16 Fr were inserted into the cardiopulmonary bypass circuit with a "Y" connector. The flow was adjusted in 100 ml/min increments within typical flow ranges for each cannula. Pre-cannula and post-cannula pressures were measured to calculate the pressure drop. RESULTS: Utilizing a pressure drop limit of 100 mmHg, our results suggest a recommended flow limit of 500, 900, 1400, 2600, and 3100 mL/min for Braile arterial cannulae sizes 8, 10, 12, 14, and 16 Fr, respectively. For Medtronic DLP arterial cannulae sizes 8, 10, 12, 14, and 16 Fr, the recommended flow limit is 600, 1100, 1700, 2700, and 3300 mL/min, respectively. CONCLUSION: This study reinforces discrepancies in pressure drop between cannulae of the same diameter supplied by different manufacturers and the importance of independent translational research to evaluate components' performance.


Assuntos
Cânula , Ponte Cardiopulmonar , Criança , Humanos , Hemodinâmica , Modelos Cardiovasculares , Desenho de Equipamento
2.
Rev. bras. cir. cardiovasc ; Rev. bras. cir. cardiovasc;39(1): e20230110, 2024. tab, graf
Artigo em Inglês | LILACS-Express | LILACS | ID: biblio-1521674

RESUMO

ABSTRACT Objective: To define a reference chart comparing pressure drop vs. flow generated by a set of arterial cannulae currently utilized in cardiopulmonary bypass conditions in pediatric surgery. Methods: Cannulae from two manufacturers were selected considering their design and outer and inner diameters. Cannula performance was evaluated in terms of pressure drop vs. flow during simulated cardiopulmonary bypass conditions. The experimental circuits consisted of a Jostra HL-20 roller pump, a Quadrox-i pediatric oxygenator (Maquet Cardiopulmonary AG, Rastatt, Germany), and a custom pediatric tubing set. The circuit was primed with lactated Ringer's solution only (first condition) and with human packed red blood cells added (second condition) to achieve a hematocrit of 30%. Cannula sizes 8 to 16 Fr were inserted into the cardiopulmonary bypass circuit with a "Y" connector. The flow was adjusted in 100 ml/min increments within typical flow ranges for each cannula. Pre-cannula and post-cannula pressures were measured to calculate the pressure drop. Results: Utilizing a pressure drop limit of 100 mmHg, our results suggest a recommended flow limit of 500, 900, 1400, 2600, and 3100 mL/min for Braile arterial cannulae sizes 8, 10, 12, 14, and 16 Fr, respectively. For Medtronic DLP arterial cannulae sizes 8, 10, 12, 14, and 16 Fr, the recommended flow limit is 600, 1100, 1700, 2700, and 3300 mL/min, respectively. Conclusion: This study reinforces discrepancies in pressure drop between cannulae of the same diameter supplied by different manufacturers and the importance of independent translational research to evaluate components' performance.

SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa