OBJECTIVE: To assess the safety of the Otologics fully implantable hearing system after 1 year of use in a Phase I clinical trial. STUDY DESIGN: Repeated-measures within-subjects design. SETTING: Procedures were performed in a variety of facilities, including a university, military, and private hospital's ambulatory surgical center and outpatient clinical audiologic test facilities. PATIENTS: Adult patients with bilateral moderate to severe sensorineural hearing loss. INTERVENTION(S): Surgical insertion of this prosthesis included an atticotomy to expose the incus, securing the transducer to the mastoid bone, attaching the transducer tip to the incus via insertion into a laser-drilled hole, and postauricular implantation of the microphone/battery/electronics capsule. MAIN OUTCOME MEASURE(S): Subjective patient benefit, aided sound field thresholds, and speech discrimination with the subject's own, appropriately fit, walk-in hearing aid(s) and the prosthesis were assessed. RESULTS: There were no pre-post-implant differences noted for bone conduction: slight differences were noted in the pre-post-implant air conduction results (p < 0.05). These differences were attributed to the healing process and reversed to almost preimplant assessment levels by the third-month evaluation. Pure-tone averages and monaural word recognition scores were slightly better for the walk-in-aided condition (p < 0.05), whereas the patient benefit scales favored the postoperative implant-aided conditions.Adverse effects of the implant were encountered on 14 occasions after the implantation of the 20 subjects. With the exception of partial device extrusions (that occurred later), all were rectified by the time of initial activation. At the 12-month data collection point, problems that had been encountered by subjects included 1) partial device extrusion (3 subjects), necessitating explantation in 2; 2) loss of external communication (2 subjects), resulting in 1 explantation; and 3) increased charging times beyond 1.5 hours (7), resulting in 3 explantations and 2 patients not using their device while awaiting explantation. CONCLUSION: Phase I trial results provide evidence that this fully implantable device can provide sound amplification to sensorineural hearing loss patients, with performance results similar to the patients' walk-in hearing aids.
Assuntos
Implantes Cocleares , Perda Auditiva Neurossensorial/cirurgia , Adulto , Idoso , Idoso de 80 Anos ou mais , Limiar Auditivo , Implante Coclear , Implantes Cocleares/efeitos adversos , Falha de Equipamento , Feminino , Seguimentos , Perda Auditiva Neurossensorial/psicologia , Humanos , Testes de Linguagem , Masculino , Pessoa de Meia-Idade , Satisfação do Paciente , Software , Percepção da Fala , Inquéritos e Questionários , Resultado do TratamentoRESUMO
OBJECTIVES: The safety of the Otologics Fully-Implantable MET Ossicular Stimulator was assessed in adult patients with bilateral moderate to severe sensorineural hearing loss. METHODS: Surgical implantation of the ossicular stimulator was performed. A repeated-measure, within-subjects design assessed safety and aided sound field thresholds and speech performances with the subject's own, appropriately fitted, walk-in hearing aid(s) and the Otologics Fully-Implantable MET Ossicular Stimulator. RESULTS: Twenty patients were implanted and activated as part of the Phase I clinical trial. Results demonstrated 10-20 dB of functional gain across audiometric frequencies. Pure tone averages and monaural word recognition scores were slightly better for the walk-in-aided condition, while patient benefit scales favored the postoperative implant-aided conditions. CONCLUSION/SIGNIFICANCE: Although monaural word scores and aided thresholds favored the walk-in-aided condition, preliminary results indicate that the Otologics MET Fully-Implantable Ossicular Stimulator is an alternative to currently available hearing aids in patients with sensorineural hearing loss.