Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 5 de 5
Filtrar
Mais filtros

Base de dados
Tipo de documento
Assunto da revista
País de afiliação
Intervalo de ano de publicação
1.
BMC Cancer ; 22(1): 673, 2022 Jun 20.
Artigo em Inglês | MEDLINE | ID: mdl-35725457

RESUMO

BACKGROUND: A novel device for supine positioning in breast radiotherapy for patients with large or pendulous breasts has been developed and tested in phase II studies. This trial is designed to assess the efficacy of the device to reduce skin toxicity and unwanted normal tissue dose in comparison to the current clinical standard for supine breast support during breast radiotherapy. METHODS: Patients at high risk for moist desquamation, having infra-mammary fold or lateral ptosis, will be randomized into two arms. Patients in the control arm will receive breast radiotherapy with supine positioning using current standard of care. Patients in the experimental arm will be positioned supine with the novel device. The primary endpoint is the incidence of moist desquamation in the infra-mammary fold. We hypothesize a 20% reduction (from 50 to 30%) in the rate of moist desquamation in the study arm versus the control arm. For 80% power, two-tailed α = 0.05 and 10% loss to follow up, 110 patients will be assigned to each arm. The proportion of patients experiencing moist desquamation in the two arms will be compared using logistic regression adjusting for brassiere cup size, skin fold size, body mass index, smoking status, and dose fractionation schedule. An unadjusted comparison will also be made using the chi-square test, or Fisher's exact test, if appropriate. Secondary endpoints include dose-volume statistics for the lung and heart, skin dose and clinical parameters including setup time, reproducibility, and staff experience with setup procedures. Patient-reported pain, skin condition interference with sleep and daily activities, and comfort during treatment are also secondary endpoints. DISCUSSION: Based on results from earlier phase II studies, it is expected that the device-enabled elimination of infra-mammary fold should reduce toxicity and improve quality of life for this patient population. Earlier studies showed reduction in dose to organs at risk including lung and heart, indicating potential for other long-term benefits for patients using the device. This study is limited to acute skin toxicity, patient-reported outcomes, and clinical factors to inform integration of the device into standard breast radiotherapy procedures. TRIAL REGISTRATION: Clinicaltrials.gov identifier: NCT04257396 . Registered February 6 2020.


Assuntos
Neoplasias da Mama , Dermatopatias , Neoplasias da Mama/etiologia , Neoplasias da Mama/radioterapia , Neoplasias da Mama/cirurgia , Fibra de Carbono , Ensaios Clínicos Fase III como Assunto , Feminino , Humanos , Mastectomia Segmentar/métodos , Estudos Multicêntricos como Assunto , Qualidade de Vida , Radioterapia Adjuvante/efeitos adversos , Ensaios Clínicos Controlados Aleatórios como Assunto , Reprodutibilidade dos Testes
2.
Adv Radiat Oncol ; 9(1): 101318, 2024 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-38260224

RESUMO

Purpose: Moist desquamation (MD) is a concerning acute side effect of radiation therapy for breast cancer, often seen in skin folds for patients having large or pendulous breasts. In vivo skin dosimetry, clinical assessments, and patient-reported skin reactions were used to determine a relationship between dose-area metrics and the development of MD, to lend insight into skin tolerances and possibly guide future treatment planning dose constraints. Methods and Materials: Skin dose was measured using GafChromic film on the inner surface of an early prototype carbon-fiber accessory for breast support to remove the inframammary fold in 20 patients at high risk of developing MD undergoing adjuvant whole breast radiation therapy. Prescribed doses were 42.5 Gray (Gy) in 16 fractions or 50 Gy in 25 fractions using 6 to 15 MV x-rays. To account for fraction size differences, analysis was performed using the equivalent dose in 2 Gy fractions using α/ß = 11 (EQD211). MD was assessed out to 2 weeks post radiation therapy by trained therapists and by a patient-reported outcome questionnaire. Results: Statistically significant differences in areas receiving 30 to 48 Gy (EQD211) were observed between patients who did and did not develop MD in the inframammary area. Patients receiving EQD211 maximum dose ≤ 46 Gy and ≥ 38 Gy to ≤ 50 cm2 of their breast skin did not develop MD. Conclusions: The findings of this study offer insight into the relationship between skin toxicity and areas of skin irradiated to doses up to 50 Gy. Potential skin dose constraints to test in future studies to prevent MD are suggested.

3.
Curr Oncol ; 29(7): 4734-4747, 2022 07 07.
Artigo em Inglês | MEDLINE | ID: mdl-35877236

RESUMO

There has been an increasing interest in patient-reported outcome (PRO) measures in both the clinical and research settings to improve the quality of life among patients and to identify when clinical intervention may be needed. The primary purpose of this prospective study was to validate an acute breast skin toxicity PRO measure across a broad sample of patient body types undergoing radiation therapy. Between August 2018 and September 2019, 134 women undergoing adjuvant breast radiotherapy (RT) consented to completing serial PRO measures both during and post-RT treatment and to having their skin assessed by trained trial radiation therapists. There was high patient compliance, with 124 patients (92.5%) returning to the clinic post-RT for at least one staff skin assessment. Rates of moist desquamation (MD) in the infra-mammary fold (IMF) by PRO were compared with skin assessments completed by trial radiation therapists. There was high sensitivity (86.5%) and good specificity (79.4%) between PRO and staff-reported presence of MD in the IMF, and there was a moderate correlation between the peak severity of the MD reported by PRO and assessed by staff (rho = 0.61, p < 0.001). This prospective study validates a new PRO measure to monitor the presence of MD in the IMF among women receiving breast RT.


Assuntos
Radiodermite , Feminino , Humanos , Mastectomia Segmentar , Medidas de Resultados Relatados pelo Paciente , Estudos Prospectivos , Qualidade de Vida , Radiodermite/tratamento farmacológico
4.
Pract Radiat Oncol ; 11(6): 470-479, 2021.
Artigo em Inglês | MEDLINE | ID: mdl-34303034

RESUMO

PURPOSE: This pilot study (ClinicalTrials.gov NCT04543851) investigates a novel breast positioning device using a low density, high tensile carbon-fiber cradle to support the breast, remove the inframammary fold, and reduce dose to organs at risk for whole breast radiation therapy in the supine position. METHODS AND MATERIALS: Thirty patients with inframammary folds ≥1 cm or lateral ptosis in supine treatment position were planned with standard positioning and with a carbon-fiber Adjustable Reusable Accessory (CARA) breast support. Twenty patients received whole breast with or without regional nodal irradiation with 42.5 Gy in 16 fractions or 50 Gy in 25 fractions using CARA. Median body mass index was 32 in this study. RESULTS: CARA removed all inframammary folds and reduced V20Gyipsilateral lung, V105%breast, and V50% body, without compromising target coverage. Median (range) V20Gyipsilateral lung for whole breast radiation therapy was 12.3% (1.4%-28.7%) with standard of care versus 10.9% (1.2%-17.3%) with CARA (Wilcoxon P = .005). Median V105% breast was 8.0% (0.0%-29%) with standard of care versus 4.0% (0.0%-23%) with CARA (P = .006) and median V50% body was 3056 mL (1476-5285 mL) versus 2780 mL (1415-5123 mL) with CARA (P = .001). CARA was compatible with deep inspiration breath hold and achieved median V25Gyheart = 0.1% (range 0%-1.9%) for all patients with left breast cancer. Skin reactions with CARA were consistent with historical data and daily variation in treatment setup was consistent with standard supine positioning. CONCLUSIONS: CARA can reduce V105%breast, lung and normal tissue dose, and remove the inframammary fold for breast patients with large or pendulous breasts and high body mass index treated in the supine position, without compromising target coverage. CARA will undergo further study in a randomized controlled trial.


Assuntos
Neoplasias da Mama , Órgãos em Risco , Neoplasias da Mama/radioterapia , Fibra de Carbono , Feminino , Coração , Humanos , Projetos Piloto , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
5.
Radiother Oncol ; 90(1): 56-9, 2009 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-18417232

RESUMO

Surgical clips were investigated for partial breast image-guided radiotherapy (IGRT). Small titanium clips were insufficiently well visualised. Medium tantalum clips were best for megavoltage IGRT and small tantalum clips were best for floor mounted kilovoltage IGRT (ExacTrac). Both small tantalum and medium titanium clips were suitable for isocentric kilovoltage IGRT.


Assuntos
Neoplasias da Mama/radioterapia , Imagens de Fantasmas , Instrumentos Cirúrgicos , Artefatos , Humanos , Modelos Logísticos , Tantálio , Titânio , Tomografia Computadorizada por Raios X
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa