Since the advent of high-throughput screening (HTS) in the early 1990s, parallel multichannel liquid handlers have become a mainstay in every drug discovery setting. Although several peer-reviewed publications have discussed methods and criteria for stamping multiwell copies, there is very little information about establishing a standard operating procedure (SOP) for standard (microliter-level) serial dilutions of compounds used in dose-response experiments. The authors discuss the 4 main criteria any serial dilution process must pass (accuracy, precision, fold dilution, and outliers) and the process for establishing thresholds for all of these values in a compound management or biological screening laboratory. The thresholds need to be both low enough to be acceptable from a biological potency variability perspective and high enough to allow the instruments to pass the quality assurance (QA) analysis on a regular basis. In this article, the authors suggest suitable thresholds arrived at by a variety of methods, including trend analysis of QA data, survey questionnaire from the main stakeholders (screening scientists, chemists), and published criteria for single-shot stamping. A mathematical analysis of the effect of threshold values on estimated XC(50)s was performed to ensure that the variability introduced by the serial dilution step is within acceptable overall variability limits.