RESUMO
BACKGROUND: In this study, we investigate the impact of increased PET acquisition time per bed position on lesion detectability, standard uptake value, and image noise in 68Ga-PSMA-11 PET/MRI scans. METHODS: Scans of twenty patients were analyzed in this study. Patients were injected with 68Ga-PSMA-11 (mean, 5.50 ± 1.49 mCi) and imaged on a 3.0 T time-of-flight PET/MRI. PET images were retrospectively reconstructed using 0.5, 1, 2, 4, 7, and 10 min of PET data. Lesion detectability was evaluated on a 5-point Likert Scale for each lesion in each reconstruction. Quantitative analysis was performed measuring image noise and lesion uptake. RESULTS: A total of 55 lesions were identified, and lesion detectability increased from 2.07 ± 1.14 for 0.5 min to 4.93 ± 0.26 for 10 min (p < 0.001), with no significant difference detected between 7 and 10 min of scan time. Average SUVmax decreased from 9.89 ± 6.62 for 0.5 min to 8.64 ± 6.81 for 10 min. Noise decreased from 0.72 ± 0.22 for 0.5 min to 0.31 ± 0.12 for 10 min (p < 0.001) and were nearly equivalent between 7 and 10 min. Pairwise interaction terms between size, SUVmax, and scan time were all found to be significant, although the interaction term between SUVmax and scan time was found to be the most significant. CONCLUSIONS: Increased acquisition duration improves image quality by increasing detectability and reducing noise. In patients with biochemical recurrence, increased acquisition time up to 7 min improves lesion detection.