IMRT delivery verification using a spiral phantom.
Richardson, Susan L; Tomé, Wolfgang A; Orton, Nigel P; McNutt, Todd R; Paliwal, Bhudatt R.
Afiliação
  • Richardson SL; Department of Radiology, Baylor College of Medicine, Houston, Texas 77030, USA.
Med Phys ; 30(9): 2553-8, 2003 Sep.
Article em En | MEDLINE | ID: mdl-14528978
ABSTRACT
In this paper we report on the testing and verification of a system for IMRT delivery quality assurance that uses a cylindrical solid water phantom with a spiral trajectory for radiographic film placement. This spiral film technique provides more complete dosimetric verification of the entire IMRT treatment than perpendicular film methods, since it samples a three-dimensional dose subspace rather than using measurements at only one or two depths. As an example, the complete analysis of the predicted and measured spiral films is described for an intracranial IMRT treatment case. The results of this analysis are compared to those of a single field perpendicular film technique that is typically used for IMRT QA. The comparison demonstrates that both methods result in a dosimetric error within a clinical tolerance of 5%, however the spiral phantom QA technique provides a more complete dosimetric verification while being less time consuming. To independently verify the dosimetry obtained with the spiral film, the same IMRT treatment was delivered to a similar phantom in which LiF thermoluminescent dosimeters were arranged along the spiral trajectory. The maximum difference between the predicted and measured TLD data for the 1.8 Gy fraction was 0.06 Gy for a TLD located in a high dose gradient region. This further validates the ability of the spiral phantom QA process to accurately verify delivery of an IMRT plan.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Radioterapia Conformacional / Dosimetria Fotográfica / Neoplasias Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans País/Região como assunto: America do norte Idioma: En Revista: Med Phys Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Imagens de Fantasmas / Radioterapia Conformacional / Dosimetria Fotográfica / Neoplasias Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans País/Região como assunto: America do norte Idioma: En Revista: Med Phys Ano de publicação: 2003 Tipo de documento: Article País de afiliação: Estados Unidos