Alternative antibody for the detection of CA19-9 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access GI Monitor assay on the UniCel Dxl 800 Immunoassay System.
Clin Chem Lab Med
; 46(5): 600-11, 2008.
Article
em En
| MEDLINE
| ID: mdl-18598202
ABSTRACT
BACKGROUND:
Gastrointestinal cancer antigen CA19-9 is known as a valuable marker for the management of patients with pancreatic cancer.METHODS:
The analytical and clinical performance of the Access GI Monitor assay (Beckman Coulter) was evaluated on the UniCel Dxl 800 Immunoassay System at five different European sites and compared with a reference method, defined as CA19-9 on the Elecsys System (Roche Diagnostics).RESULTS:
Total imprecision (%CV) of the GI Monitor ranged between 3.4% and 7.7%, and inter-laboratory reproducibility between 3.6% and 4.0%. Linearity upon dilution showed a mean recovery of 97.4% (SD + 7.2%). Endogenous interferents had no influence on GI Monitor levels (mean recoveries hemoglobin 103%, bilirubin 106%, triglycerides 106%). There was no high-dose hook effect up to 115,000 kU/L. Clinical performance investigated in sera from 1811 individuals showed a good correlation between the Access GI Monitor and Elecsys CA19-9 (R = 0.959, slope = 1.004, intercept = +0.17). GI Monitor serum levels were low in healthy individuals (n = 267, median = 6.0 kU/L, 95th percentile=23.1 kU/L), higher in individuals with various benign diseases (n = 550, medians = 5.8-13.4 kU/L, 95th percentiles = 30.1-195.5 kU/L) and even higher in individuals suffering from various cancers (n = 995, medians = 8.4-233.8 kU/L, 95th percentiles = 53.7-13,902 kU/L). Optimal diagnostic accuracy for cancer detection against the relevant benign control group by the GI Monitor was found for pancreatic cancer [area under the curve (AUC) 0.83]. Results for the reference CA19-9 assay were comparable (AUC 0.85).CONCLUSIONS:
The Access GI Monitor provides very good methodological characteristics and demonstrates an excellent analytical and clinical correlation with the Elecsys CA19-9. The GI Monitor shows the best diagnostic accuracy in pancreatic cancer. Our results also suggest a clinical value of the GI Monitor in other cancers.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Neoplasias Pancreáticas
/
Imunoensaio
/
Antígeno CA-19-9
/
Anticorpos
Tipo de estudo:
Clinical_trials
/
Diagnostic_studies
/
Evaluation_studies
/
Prognostic_studies
Limite:
Adolescent
/
Adult
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
País/Região como assunto:
Europa
Idioma:
En
Revista:
Clin Chem Lab Med
Assunto da revista:
QUIMICA CLINICA
/
TECNICAS E PROCEDIMENTOS DE LABORATORIO
Ano de publicação:
2008
Tipo de documento:
Article
País de afiliação:
Alemanha