ABSTRACT
An episode of device-related proarrhythmia is reported. Ventricular tachycardia was induced by ventricular pacing in a dual-chamber implantable cardioverter-defibrillator due to an inappropriate interaction of programmed device settings with the atrial preference pacing, an automatic algorithm designed to prevent atrial tachyarrhythmias.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Terapia Assistida por Computador / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Prognostic_studies Limite: Aged / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Terapia Assistida por Computador / Taquicardia Ventricular / Desfibriladores Implantáveis Tipo de estudo: Prognostic_studies Limite: Aged / Humans / Male Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Espanha