ICD proarrhythmia as a consequence of an interaction with an algorithm to prevent atrial arrhythmias.
Pacing Clin Electrophysiol
; 32(8): 1096-8, 2009 Aug.
Article
em En
| MEDLINE
| ID: mdl-19659632
ABSTRACT
An episode of device-related proarrhythmia is reported. Ventricular tachycardia was induced by ventricular pacing in a dual-chamber implantable cardioverter-defibrillator due to an inappropriate interaction of programmed device settings with the atrial preference pacing, an automatic algorithm designed to prevent atrial tachyarrhythmias.