Your browser doesn't support javascript.
loading
Failure of larger-diameter metal-on-metal total hip arthroplasty resulting from anterior iliopsoas impingement.
Browne, James A; Polga, David J; Sierra, Rafael J; Trousdale, Robert T; Cabanela, Miguel E.
Afiliação
  • Browne JA; Department of Orthopaedics, University of Virginia, Charlottesville, VA 22908, USA.
J Arthroplasty ; 26(6): 978.e5-8, 2011 Sep.
Article em En | MEDLINE | ID: mdl-21130601
ABSTRACT
Anterior iliopsoas impingement is a recognized cause of persistent groin pain after total hip arthroplasty. We report 3 patients with failed total hip arthroplasties resulting from anterior iliopsoas and capsular impingement secondary to a metal femoral ball with a diameter larger than the native femoral head. All patients had the same implant design. Resolution of symptoms occurred in all patients after revision surgery. To our knowledge, this is the first report of this potential failure mechanism with these large-diameter implant designs and should be considered in the appropriate clinical scenario.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Músculos Psoas / Artroplastia de Quadril / Impacto Femoroacetabular / Prótese de Quadril / Metais Tipo de estudo: Etiology_studies Limite: Aged / Female / Humans / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Falha de Prótese / Músculos Psoas / Artroplastia de Quadril / Impacto Femoroacetabular / Prótese de Quadril / Metais Tipo de estudo: Etiology_studies Limite: Aged / Female / Humans / Middle aged Idioma: En Revista: J Arthroplasty Assunto da revista: ORTOPEDIA Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos