Detection of ground-glass opacities by use of hybrid iterative reconstruction (iDose) and low-dose 256-section computed tomography: a phantom study.
Higuchi, Ko; Nagao, Michinobu; Matsuo, Yoshio; Sunami, Shunya; Kamitani, Takeshi; Jinnouchi, Mikako; Yonezawa, Masato; Yamasaki, Yuzo; Yabuuchi, Hidetake; Hatkenaka, Masamitsu; Honda, Hiroshi.
Afiliação
  • Higuchi K; Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-Ku, Fukuoka-city, Fukuoka 812-8582, Japan. khiguchi@radiol.med.kyushu-u.ac.jp
Radiol Phys Technol ; 6(2): 299-304, 2013 Jul.
Article em En | MEDLINE | ID: mdl-23400447
ABSTRACT
The detection of ground-glass opacities (GGOs) is an important issue in lung cancer screening with low-dose CT. The iterative reconstruction (IR) technique has the ability to improve the image quality relative to the filtered back projection (FBP) technique with low-dose CT. Our purpose was to investigate the ability to detect GGO in a chest phantom using a low-dose CT and hybrid IR, named iDose. Simulated GGOs in a chest phantom were scanned with 256-section CT at tube current second products of 20, 50, 100, and 200 mAs. Five radiologists visually assessed the detectability of GGOs in the phantom. The contrast-to-noise ratio (CNR) for GGOs was used as an estimate of image quality. Comparison of the detectability and CNR between standard images with 200 mAs-FBP and low-dose images with 20, 50, and 100-mAs FBP/iDose were performed by ANOVA with Dunnett's and Tukey's test. The detectability was significantly lower at 20-mAs FBP/iDose and 50-mAs FBP than that at 200-mAs FBP (p < 0.05). There was no significant difference between 50-mAs iDose and 200-mAs FBP and between 100-mAs iDose/FBP and 200-mAs FBP. The CNR was significantly higher on iDose images than that on FBP images at each mAs value. The CNR at 200-mAs FBP was the same as that at 50-mAs iDose (CNR1.8). The hybrid IR technique and low-dose CT imaging with 50 mAs enabled noise and to maintain the detectability for GGOs in a chest phantom that is equivalent to the reference acquisitions of 200 mAs with FBP.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Radiografia Torácica / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Nódulo Pulmonar Solitário / Imagens de Fantasmas / Vidro Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Radiol Phys Technol Assunto da revista: BIOFISICA / RADIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Algoritmos / Radiografia Torácica / Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Nódulo Pulmonar Solitário / Imagens de Fantasmas / Vidro Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Radiol Phys Technol Assunto da revista: BIOFISICA / RADIOLOGIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Japão