Notching of the femoral stem neck in metal-on-metal total hip replacement: a case report.
Leung, Ka-Hei; Chiu, Kwong-Yuen; Ng, Fu-Yuen; Yin, Li.
Afiliação
  • Leung KH; Department of Orthopaedics and Traumatology, Queen Mary Hospital, Hong Kong. leungkahei@hotmail.com
J Orthop Surg (Hong Kong) ; 21(1): 113-6, 2013 Apr.
Article em En | MEDLINE | ID: mdl-23630003
We report a case of impingement in a metal-on-metal total hip replacement causing both notching of the femoral stem neck and aseptic loosening of the acetabular component. The acetabular component was inserted in excessive anteversion. The femoral stem and acetabular components were retrieved. A larger femoral head was used, and an acetabular component was placed in a less anteverted position. Intra-operative testing through different ranges of movement is recommended to identify potential impingement.
Assuntos
Palavras-chave
Buscar no Google
Base de dados: MEDLINE Assunto principal: Colo do Fêmur / Impacto Femoroacetabular / Próteses Articulares Metal-Metal / Prótese de Quadril Tipo de estudo: Prognostic_studies Limite: Female / Humans / Middle aged Idioma: En Revista: J Orthop Surg (Hong Kong) Assunto da revista: ORTOPEDIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Hong Kong
Buscar no Google
Base de dados: MEDLINE Assunto principal: Colo do Fêmur / Impacto Femoroacetabular / Próteses Articulares Metal-Metal / Prótese de Quadril Tipo de estudo: Prognostic_studies Limite: Female / Humans / Middle aged Idioma: En Revista: J Orthop Surg (Hong Kong) Assunto da revista: ORTOPEDIA Ano de publicação: 2013 Tipo de documento: Article País de afiliação: Hong Kong