Your browser doesn't support javascript.
loading
Evaluation of the analytical performance of the novel NS-Prime system and examination of temperature stability of fecal transferrin compared with fecal hemoglobin as biomarkers in a colon cancer screening program.
Demian, Wael L L; Collins, Stacy; Fowler, Candace; McGrath, Jerry; Antle, Scott; Moores, Zoë; Hollohan, Deborah; Lacey, Suzanne; Banoub, Joseph; Randell, Edward.
Afiliação
  • Demian WLL; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • Collins S; Department of Biochemistry, Memorial University of Newfoundland, 232 Elizabeth Avenue, St. John's, Newfoundland, Canada A1B 3X9.
  • Fowler C; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • McGrath J; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • Antle S; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • Moores Z; Faculty of Medicine, Memorial University of Newfoundland, St. John's, Newfoundland, Canada A1B 3V6.
  • Hollohan D; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • Lacey S; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • Banoub J; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
  • Randell E; Eastern Health Authority, St. John's, Newfoundland, Canada A1B 3V6.
Pract Lab Med ; 2: 29-36, 2015 Aug 01.
Article em En | MEDLINE | ID: mdl-28932802
ABSTRACT

OBJECTIVE:

To examine the analytical aspects of fecal transferrin (Tf) and hemoglobin (Hb) measured on the NS-Prime analyzer for use in a colon cancer screening program. DESIGNS AND

METHODS:

Method evaluation and temperature stability studies for fecal Tf and Hb were completed. A method comparison was carried out against the NS-Plus system using samples collected from 254 screening program participants. A further 200 samples were analyzed to help determine suitable reference limits for fecal Tf using these systems.

RESULTS:

The assay for fecal Tf showed acceptable linearity, precision, and recovery, and showed minimal carryover with low potential for impact by the prozone effect. The 95th percentile for fecal Tf obtained for the reference population was 4.9 µg/g feces. The collection device sufficiently maintained fecal Tf and Hb stability for at least 7 days at room temperature, 4 °C, and -20 °C. Fecal Tf and Hb were most stable at 4 °C and -20 °C, but showed considerable loss (20-40%) of both proteins at 37 °C within the first 7 days. Mixing small amounts of blood into diluted fecal samples maintained at 37 °C for various time periods showed >50% loss of both proteins within 1 h of incubation.

CONCLUSIONS:

The NS-Prime analyzer showed acceptable performance for fecal Tf and Hb. These studies suggest that use of both Tf and Hb together as biomarkers will result in higher positivity rates, but this may not be attributed to greater stability of Tf over Hb in human feces.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Evaluation_studies / Screening_studies Idioma: En Revista: Pract Lab Med Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Evaluation_studies / Screening_studies Idioma: En Revista: Pract Lab Med Ano de publicação: 2015 Tipo de documento: Article