Contrast agent concentration optimization in CTA using low tube voltage and dual-energy CT in multiple vendors: a phantom study.
van Hamersvelt, Robbert W; Eijsvoogel, Nienke G; Mihl, Casper; de Jong, Pim A; Schilham, Arnold M R; Buls, Nico; Das, Marco; Leiner, Tim; Willemink, Martin J.
Afiliação
  • van Hamersvelt RW; Department of Radiology, University Medical Center Utrecht, Utrecht University, P. O. Box 85500, 3508 GA, Utrecht, The Netherlands. R.W.vanHamersvelt-3@umcutrecht.nl.
  • Eijsvoogel NG; Department of Radiology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Mihl C; Department of Radiology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • de Jong PA; Department of Radiology, University Medical Center Utrecht, Utrecht University, P. O. Box 85500, 3508 GA, Utrecht, The Netherlands.
  • Schilham AMR; Department of Radiology, University Medical Center Utrecht, Utrecht University, P. O. Box 85500, 3508 GA, Utrecht, The Netherlands.
  • Buls N; Radiology, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel (UZ Brussel), Brussels, Belgium.
  • Das M; Department of Radiology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Leiner T; Department of Radiology, University Medical Center Utrecht, Utrecht University, P. O. Box 85500, 3508 GA, Utrecht, The Netherlands.
  • Willemink MJ; Department of Radiology, University Medical Center Utrecht, Utrecht University, P. O. Box 85500, 3508 GA, Utrecht, The Netherlands.
Int J Cardiovasc Imaging ; 34(8): 1265-1275, 2018 Aug.
Article em En | MEDLINE | ID: mdl-29516228
ABSTRACT
We investigated the feasibility and extent to which iodine concentration can be reduced in computed tomography angiography imaging of the aorta and coronary arteries using low tube voltage and virtual monochromatic imaging of 3 major dual-energy CT (DECT) vendors. A circulation phantom was imaged with dual source CT (DSCT), gemstone spectral imaging (GSI) and dual-layer spectral detector CT (SDCT). For each scanner, a reference scan was acquired at 120 kVp using routine iodine concentration (300 mg I/ml). Subsequently, scans were acquired at lowest possible tube potential (70, 80, 80 kVp, respectively), and DECT-mode (80/150Sn, 80/140 and 120 kVp, respectively) in arterial phase after administration of iodine (300, 240, 180, 120, 60, 30 mg I/ml). Objective image quality was evaluated using attenuation, CNR and dose corrected CNR (DCCNR) measured in the aorta and left main coronary artery. Average DCCNR at reference was 227.0, 39.7 and 60.2 for DSCT, GSI and SDCT. Maximum iodine concentration reduction without loss of DCCNR was feasible down to 180 mg I/ml (40% reduced) for DSCT (DCCNR 467.1) and GSI (DCCNR 46.1) using conventional CT low kVp, and 120 mg I/ml (60% reduced) for SDCT (DCCNR 171.5) using DECT mode. Low kVp scanning and DECT allows for 40-60% iodine reduction without loss in image quality compared to reference. Optimal scan protocol and to which extent varies per vendor. Further patient studies are needed to extend and translate our findings to clinical practice.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta / Iohexol / Imagens de Fantasmas / Meios de Contraste / Vasos Coronários / Angiografia por Tomografia Computadorizada Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Int J Cardiovasc Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aorta / Iohexol / Imagens de Fantasmas / Meios de Contraste / Vasos Coronários / Angiografia por Tomografia Computadorizada Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Int J Cardiovasc Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Holanda