Your browser doesn't support javascript.
loading
Intraoperative Blood Pressure Monitoring in Obese Patients.
Anesthesiology ; 134(2): 179-188, 2021 02 01.
Article em En | MEDLINE | ID: mdl-33326001
BACKGROUND: The optimal method for blood pressure monitoring in obese surgical patients remains unknown. Arterial catheters can cause potential complications, and noninvasive oscillometry provides only intermittent values. Finger cuff methods allow continuous noninvasive monitoring. The authors tested the hypothesis that the agreement between finger cuff and intraarterial measurements is better than the agreement between oscillometric and intraarterial measurements. METHODS: This prospective study compared intraarterial (reference method), finger cuff, and oscillometric (upper arm, forearm, and lower leg) blood pressure measurements in 90 obese patients having bariatric surgery using Bland-Altman analysis, four-quadrant plot and concordance analysis (to assess the ability of monitoring methods to follow blood pressure changes), and error grid analysis (to describe the clinical relevance of measurement differences). RESULTS: The difference (mean ± SD) between finger cuff and intraarterial measurements was -1 mmHg (± 11 mmHg) for mean arterial pressure, -7 mmHg (± 14 mmHg) for systolic blood pressure, and 0 mmHg (± 11 mmHg) for diastolic blood pressure. Concordance between changes in finger cuff and intraarterial measurements was 88% (mean arterial pressure), 85% (systolic blood pressure), and 81% (diastolic blood pressure). In error grid analysis comparing finger cuff and intraarterial measurements, the proportions of measurements in risk zones A to E were 77.1%, 21.6%, 0.9%, 0.4%, and 0.0% for mean arterial pressure, respectively, and 89.5%, 9.8%, 0.2%, 0.4%, and 0.2%, respectively, for systolic blood pressure. For mean arterial pressure and diastolic blood pressure, absolute agreement and trending agreement between finger cuff and intraarterial measurements were better than between oscillometric (at each of the three measurement sites) and intraarterial measurements. Forearm performed better than upper arm and lower leg monitoring with regard to absolute agreement and trending agreement with intraarterial monitoring. CONCLUSIONS: The agreement between finger cuff and intraarterial measurements was better than the agreement between oscillometric and intraarterial measurements for mean arterial pressure and diastolic blood pressure in obese patients during surgery. Forearm oscillometry exhibits better measurement performance than upper arm or lower leg oscillometry.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Monitorização Intraoperatória / Cirurgia Bariátrica / Obesidade Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Anesthesiology Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Determinação da Pressão Arterial / Monitorização Intraoperatória / Cirurgia Bariátrica / Obesidade Tipo de estudo: Observational_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Anesthesiology Ano de publicação: 2021 Tipo de documento: Article