Suction mask device: a simple, inexpensive, and effective method of reducing spread of aerosolized particles during endoscopic endonasal surgery in the era of COVID-19.
Hara, Takuma; Zachariah, Marcus A; Li, Ruichun; Martinez-Perez, Rafael; Carrau, Ricardo L; Prevedello, Daniel M.
Afiliação
  • Hara T; Departments of1Neurosurgery and.
  • Zachariah MA; 4Department of Neurosurgery, Faculty of Medicine, University of Tsukuba, Ibaraki, Japan.
  • Li R; 3Department of Neurosurgery, University of Mississippi Medical Center, Jackson, Mississippi; and.
  • Martinez-Perez R; Departments of1Neurosurgery and.
  • Carrau RL; Departments of1Neurosurgery and.
  • Prevedello DM; Departments of1Neurosurgery and.
J Neurosurg ; 135(5): 1328-1334, 2021 Apr 23.
Article em En | MEDLINE | ID: mdl-33892478
ABSTRACT

OBJECTIVE:

Aerosol-generating procedures, including endoscopic endonasal surgery (EES), are a major risk for physicians during the COVID-19 pandemic. Techniques for reducing aerosolization and risk of transmission of COVID-19 during these procedures would be valuable to the neurosurgical community. The authors aimed to simulate the generation of small-particle aerosols during EES and craniectomy in order to develop methods to reduce the spread of aerosolized particles, and to test the effectiveness of these methods.

METHODS:

This study was performed at the Anatomical Laboratory for Visuospatial Innovations in Otolaryngology and Neurosurgery at The Ohio State University. The following two scenarios were used to measure three different particle sizes (0.3, 2.5, and 10 µm) generated 1) drilling frontotemporal bone, simulating a craniectomy; and 2) drilling sphenoid bone, simulating an endonasal approach. A suction mask device was created with the aim of reducing particle release. The presence of particles was measured without suction, with a single Frazier tip suction in the field, and with the suction mask device in addition to the Frazier suction tip. Particles were measured 12 cm from the craniectomy or endonasal drilling region.

RESULTS:

In the absence of any aerosol-reducing devices, the number of particles measured during craniectomy was significantly higher than that generated by endonasal drilling. This was true regardless of the particle size measured (0.3 µm, p < 0.001; 2.5 µm, p < 0.001; and 10 µm, p < 0.001). The suction mask device reduced the release of particles of all sizes measured in the craniectomy simulation (0.3 µm, p < 0.001; 2.5 µm, p < 0.001; and 10 µm, p < 0.001) and particles of 0.3 µm and 2.5 µm in the single Frazier suction simulation (0.3 µm, p = 0.031; and 2.5 µm, p = 0.026). The suction mask device further reduced the release of particles of all sizes during EES simulation (0.3 µm, p < 0.001; and 2.5 µm, p < 0.001) and particles of 0.3 µm and 2.5 µm in the single Frazier suction simulation (0.3 µm, p = 0.033; and 2.5 µm, p = 0.048). Large particles (10 µm) were not detected during EES.

CONCLUSIONS:

The suction mask device is a simple and effective means of reducing aerosol release during EES, and it could potentially be used during mastoidectomies. This could be a valuable tool to reduce the risk of procedure-associated viral transmission during the COVID-19 pandemic.