Your browser doesn't support javascript.
loading
Feasibility investigation of logarithmic Nakagami parametric imaging in recovering underestimated perfusion metrics of DCEUS in the uneven acoustic field.
Wang, Diya; Ma, Quanlong; Wang, Qiao; Fan, Yan; Lu, Shukuan; Su, Qiang; Zhong, Hui; Wan, Mingxi.
Afiliação
  • Wang D; Key Laboratory of Biomedical Information Engineering of Ministry of Education, School of Life Science and Technology, Xi' an Jiaotong University, Xi' an, China.
  • Ma Q; Key Laboratory of Biomedical Information Engineering of Ministry of Education, School of Life Science and Technology, Xi' an Jiaotong University, Xi' an, China.
  • Wang Q; Key Laboratory of Biomedical Information Engineering of Ministry of Education, School of Life Science and Technology, Xi' an Jiaotong University, Xi' an, China.
  • Fan Y; Alliance Franco-Chinoise, Montreal, Canada.
  • Lu S; Key Laboratory of Biomedical Information Engineering of Ministry of Education, School of Life Science and Technology, Xi' an Jiaotong University, Xi' an, China.
  • Su Q; Department of Oncology, Beijing Friendship Hospital Captial Medical Unviersity, Beijing, China.
  • Zhong H; Key Laboratory of Biomedical Information Engineering of Ministry of Education, School of Life Science and Technology, Xi' an Jiaotong University, Xi' an, China.
  • Wan M; Key Laboratory of Biomedical Information Engineering of Ministry of Education, School of Life Science and Technology, Xi' an Jiaotong University, Xi' an, China.
Med Phys ; 49(4): 2452-2461, 2022 Apr.
Article em En | MEDLINE | ID: mdl-35137426
ABSTRACT

PURPOSE:

Owing to acoustic-pressure dependence, amplitudes of backscattered-echoes of encapsulated microbubbles (MBs) are unavoidably regulated by an uneven acoustic field, resulting in the misestimation of hemodynamics in conventional amplitude-coding dynamic contrast-enhanced ultrasound (DCEUS) with focused pulse transmission. This study aimed to investigate the feasibility and performance of Nakagami statistical-feature parametric imaging to recover the above misestimation.

METHODS:

Logarithmic Nakagami parameter (m)-coding DCEUS scheme was investigated via simulation and in vitro MB phantoms as well as in vivo kidney-perfusion experiments of four rabbits in the uneven acoustic fields with two different focal depths. In vivo tissue artifacts for m estimation were suppressed by pulse-inversion second-harmonic imaging and its robustness was enhanced by multiscale moment-estimation strategy. Time-Nakagami-m curves and the corresponding perfusion metrics of intensity and volume were calculated from the logarithmic m-coding DCEUS images within the prefocal and focal regions. These curves and metrics were further compared with the perfusion curves and metrics estimated from the conventional amplitude-coding images within the same regions.

RESULTS:

Compared with amplitudes of nonlinear scattering MB echoes, their logarithmic m values were relatively independent of the changes in acoustics pressures. Compared with the fixed-scale moment-estimation, the perfusion intensity estimated from logarithmic m-coding DCEUS scheme using multiscale statistical moment-estimation had smaller differences between the prefocal and focal regions. The differences of perfusion intensity induced by an uneven acoustic field decreased to 3.47% ± 1.58 %. The differences decreased by the logarithmic m-coding DCEUS scheme were further regulated by threshold values of m estimation.

CONCLUSIONS:

The logarithmic m-coding DCEUS scheme could recover the underestimated MB backscattered-echoes and the misestimated perfusion intensity induced by the uneven acoustic field. The scheme had the potential to weaken the limitation of microvasculature identification and hemodynamic characterization marked by MBs within tissues or tumors in the uneven acoustic field.