Comparison of four different immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry assay for serum folate.
Jin, Lizi; Lu, Youli; Yi, Xilian; Zhang, Meiwei; Zhang, Jiangtao; Zhou, Weiyan; Zeng, Jie; Zhang, Tianjiao; Zhang, Chuanbao.
Afiliação
  • Jin L; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing, P.R. China.
  • Lu Y; Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
  • Yi X; Central Laboratory, Shanghai Xuhui Central Hospital/Zhongshan-Xuhui Hospital, Fudan University, Shanghai, P.R. China.
  • Zhang M; Shanghai Engineering Research Center of Phase I Clinical Research & Quality Consistency Evaluation for Drugs, Shanghai, P.R. China.
  • Zhang J; Shanghai Institute of Clinical Mass Spectrometry, Shanghai, P.R. China.
  • Zhou W; National Center for Clinical Laboratories, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing, P.R. China.
  • Zeng J; Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
  • Zhang T; Central Laboratory, Shanghai Xuhui Central Hospital/Zhongshan-Xuhui Hospital, Fudan University, Shanghai, P.R. China.
  • Zhang C; Shanghai Engineering Research Center of Phase I Clinical Research & Quality Consistency Evaluation for Drugs, Shanghai, P.R. China.
Clin Chem Lab Med ; 60(9): 1393-1402, 2022 08 26.
Article em En | MEDLINE | ID: mdl-35675883
OBJECTIVES: Accurate measurement of serum folate is essential for the diagnosis and management of various disorders. This study aims to investigate the between-method differences of four immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method. METHODS: Roche Cobas (USA), Abbott Alinity i2000 (USA), Beckman Coulter Access (USA), Mindray CL-6000i (China), and the ID-LC-MS/MS method were compared using 46 human serum samples. The results were analysed by Passing-Bablok regressions and Bland-Altman plots. A bias of 13.31% based on biological variation was used as the bias criterion. RESULTS: All the within-run and total coefficients of variation (CVs) met the specification. The folate concentrations determined by all the assays were significantly different (p=0.0028). All assays had correlation coefficients over 0.97 with each other. The 95% confidence intervals (CIs) for the slope seldom contained 1 and few 95% CIs for the intercept contained 0 in the regression equations. Compared to ID-LC-MS/MS, the biases of all assays ranged from -20.91 to 13.56 nmol/L, and the mean relative biases ranged from -9.85 to 40.33%. The predicted mean relative biases at the medical decision levels rarely met the criterion. CONCLUSIONS: Assays for serum folate had good correlations with each other but lacked good agreement. The accuracy and consistency of assays for serum folate should be measured and assessed routinely. Standardization work to improve the accuracy of serum folate assays, such as the extension of traceability to reference methods or materials, calibration standardization efforts, and assay-adjusted cut-offs should be promoted.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espectrometria de Massas em Tandem / Isótopos Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Clin Chem Lab Med Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espectrometria de Massas em Tandem / Isótopos Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Clin Chem Lab Med Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2022 Tipo de documento: Article